Results for GPC1938_X_018_A_H2


Basic Information for GPC1938_X_018_A_H2


Brief description of DB components associated to this Hybrid Assembly and associated tests


Click for packaging sticker
MetadataValue
Alt Identifier UCSC-X-0047
Serial Number 20USBHX2000778
Assembled True
Completed True
Current Location SCIPP, UC Santa Cruz
Manufacture ID GPC1938_X_018_A_H2
User Matthew Gignac

Children

ComponentTypeTimestamp
HCC Chip HCCSTAR 2022-02-20T17:10:47.422Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
ABC Chip ABCSTAR 2021-10-19T18:11:45.913Z
STAR Hybrid Flex X 2021-10-19T18:11:48.106Z
HCC Chip HCCSTARV1 2021-10-19T18:11:45.913Z
ABC Chip ABCSTARV1 2021-10-29T21:38:46.952Z
ABC Chip ABCSTARV1 2021-10-29T21:38:49.210Z
ABC Chip ABCSTARV1 2021-10-29T21:38:51.507Z
ABC Chip ABCSTARV1 2021-10-29T21:38:54.695Z
ABC Chip ABCSTARV1 2021-10-29T21:38:56.713Z
ABC Chip ABCSTARV1 2021-10-29T21:38:59.144Z
ABC Chip ABCSTARV1 2021-10-29T21:39:01.766Z
ABC Chip ABCSTARV1 2021-10-29T21:39:05.105Z
ABC Chip ABCSTARV1 2021-10-29T21:39:06.484Z
ABC Chip ABCSTARV1 2021-10-29T21:39:08.346Z

Parents

ComponentTypeTimestamp
Module 2021-12-09T19:48:53.198Z

NameCurrent StageTimestamp

Test NamePassed?Date
Glue weight (ASICs on hybrid) 2021-10-20T17:33:53.553Z
Visual Inspection (Reception) 2021-10-20T17:33:53.553Z
ASIC Metrology 2021-10-20T18:39:55.503Z
Noise Occupancy PPA
Strobe Delay PPA 2021-12-04T04:01:57.000Z
Response Curve PPA 2021-12-04T04:08:44.000Z
Pedestal Trim PPA 2021-12-04T04:01:45.000Z
Visual Inspection 2021-11-10T23:22:39.575Z
Wire bonding 2021-11-10T23:22:39.575Z

Glue weight for GPC1938_X_018_A_H2


Shown below are results for the glue weight measurements

IService::UPLOADDB                         INFO                Working on test code ASIC_GLUE_WEIGHT
GlueWeightSvc::uploadDB                    INFO                Formatting database test for ASIC_GLUE_WEIGHT
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION_RECEPTION
IService::log                              INFO                Logging status 1111 for component GPC1938_X_018_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC1938_X_018_A_H2/ASICGlueWeight_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed DB_COMMENTS.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_ASIC_GLUE_WEIGHT.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION_RECEPTION.json...
IService::install                          INFO                Installed GPC1938_X_018_A_H2.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//ASICGlueWeight_v1//GPC1938_X_015_B_H3.config
IService::next                             INFO                Scheduling GPC1938_X_015_B_H3.json file to run...
IService::next                             INFO                Writing results to /home/data/output//ASICGlueWeight_v1//GPC1938_X_015_B_H3/1/

MetadataValue
Component type ASIC
Operator Kirsten Affolder
Date 2021-10-20T17:33:53.553+00:00
Program ASIC_X
Tip size 23.000
Pressure 25.000
Dispense time 2.650
FieldWeight/Volume
Assummed density of glue 1.077
Extracted glue weight 0.045
Total volume of ASIC glue [uL] 41.790
Volume per glue dot [nL] 803.655
Weight of glue under ASICs [mg] 45.000
Weight per 5 glue dots [mg] 4.327
Weight per glue dot [mg] 0.865
Weights of ASICs [g] 0.397
Weights of ASICs plus HCC [g] 0.410
Weights of bare hybrid [g] 2.017
Weights of stuffed hybrid [g] 2.472

Metrology results


Shown below are the results of the module metrology scans

MetrologySvc::analyzeHybridThickness       INFO                Total stack up 774.555750, chip thickness 300.000000 and glue 96.534375, giving hybrid 378.021375 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 775.520500, chip thickness 300.000000 and glue 95.948071, giving hybrid 379.572429 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 776.035000, chip thickness 300.000000 and glue 91.392929, giving hybrid 384.642071 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 776.734250, chip thickness 300.000000 and glue 87.711179, giving hybrid 389.023071 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 776.814750, chip thickness 300.000000 and glue 87.205821, giving hybrid 389.608929 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 776.384750, chip thickness 300.000000 and glue 89.437357, giving hybrid 386.947393 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 777.407250, chip thickness 300.000000 and glue 92.303536, giving hybrid 385.103714 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 776.385250, chip thickness 300.000000 and glue 84.898607, giving hybrid 391.486643 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 774.493750, chip thickness 300.000000 and glue 80.783321, giving hybrid 393.710429 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 771.476000, chip thickness 300.000000 and glue 76.227536, giving hybrid 395.248464 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 772.648000, chip thickness 300.000000 and glue 86.849929, giving hybrid 385.798071 thickness
IService::UPLOADDB                         INFO                Working on test code ASIC_METROLOGY
MetrologySvc::uploadDB                     INFO                Doing DB upload...
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyHybrid_v10//GPC1938_X_018_A_H2/1//Data__GPC1938_X_018_A_H2.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyHybrid_v10//GPC1938_X_018_A_H2/1//Results__GPC1938_X_018_A_H2.txt
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyHybrid_v10//GPC1938_X_018_A_H2/1//Data__GPC1938_X_018_A_H2.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyHybrid_v10//GPC1938_X_018_A_H2/1//Results__GPC1938_X_018_A_H2.txt
IService::log                              INFO                Logging status 1111 for component GPC1938_X_018_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC1938_X_018_A_H2/MetrologyHybrid_v10-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed Summary.pdf...
IService::install                          INFO                Installed TotalStackUp.pdf...
IService::install                          INFO                Installed HybridThickness.pdf...
IService::install                          INFO                Installed Positions.pdf...
IService::install                          INFO                Installed Tilt.pdf...
IService::install                          INFO                Installed Data__GPC1938_X_018_A_H2.txt...
IService::install                          INFO                Installed Results__GPC1938_X_018_A_H2.txt...
IService::install                          INFO                Installed Results.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_ASIC_METROLOGY.json...
IService::install                          INFO                Installed GPC1938_X_018_A_H2.root...
IService::install                          INFO                Installed status.json...
TopControl::monitor                        INFO                Serivce MetrologyHybrid_v10-beta has 5 files to process! Proceeding...
MetaData::readField                        INFO                Reading file /home/data/upload//MetrologyHybrid_v10-beta//GPC1938_X_015_B_H2.config
IService::next                             INFO                Scheduling GPC1938_X_015_B_H2.txt file to run...
IService::next                             INFO                Writing results to /home/data/output//MetrologyHybrid_v10-beta//GPC1938_X_015_B_H2/1/
MetadataValue
Component type X
Operator Kirsten Affolder
Date 2021-10-20T18:39:55.503+00:00
Program version v10
Instrument SmartScope Flash 250

The thickness of the glue \(t_{glue}\) is calculated as: $$t_{glue} = \frac{1}{N_{c}}\sum_{c=1}^{N_{c}} z_{c} - \frac{1}{N_{b}}\sum_{b=1}^{N_{b}} z_{b} - t_{chip} $$ where \(t_{chip}\) is the thickness of the chip \(z_{i}\) is the z-height measurement, \(N_{i}\) is the number of measurements on a given surface, and \(c\) and \(b\) run over measurements taken on the chip and base, respectively. The assumed thickness of the chip is \(t_{chip} = 300\mu m\).

Uncertainties on the glue height measurement are estimated from the spread in z-height measurements on a given surface. On a given surface, the height uncertainty \(\sigma_{i}\) is calculated as: $$\sigma_{i} = \frac{\Delta z}{2\sqrt{N_{i}}}$$ where \(\Delta z = z_{max} - z_{min}\) and \(z_{max}\) and \(z_{min}\) are the maximum and minimum z-height measurements, respectively. In addition, a conservative \(\sigma_{chip} = 10\mu m\) uncertainty on the chip thickeness is assumed. The total uncertainty is calculated as: $$\sigma_{glue} = \sqrt{ \sigma_{c}^2 + \sigma_{b}^2 + \sigma_{chip} }$$ The glue thickness measurements for each ASIC are shown in the figure and table below.


The thickness total stack up is measured from the top of the each chip to the jig, and accounts for the combined thickness of hybrid flex, glue, and ABC/HCC chips in the case of hybrid.


The positions of the ASICs are measured on the 95x95 \(\mu m^2\) square pads on the opposite side of the sensor wire bonds. These measured positions are compared against the technical drawns to calculated the relative ASIC position. In some cases, the SmartScope incorrectly measures a different object than these square pads, and using properties (such as the perimeter), the position is corrected. The positions are shown in the figure below. If any corrections are applied, they are shown as transparent points.

The hybrid thickness is plotted below. It is calculated from the total stack up height minus the assembed chip thickness and measured glue thickness.

The tilt of a chip can be calculated in either the direction along the surface of a chip. We define the x-direction to be any points whose difference is much larger along x, and similarly for y. The tilt in both cases is defined to be: $$tilt = \frac{|\Delta z|}{\sqrt{\Delta x^{2} + \Delta y^{2}}} $$

Wirebonding for GPC1938_X_018_A_H2


Wirebonding summaries will be given below

WireBondSvc::run                           INFO                Loading group FAILED_ASIC_BACKEND
WireBondSvc::run                           INFO                Loading group FAILED_HYBRID_TO_PANEL
WireBondSvc::run                           INFO                Loading group REPAIRED_ASIC_BACKEND
WireBondSvc::run                           INFO                Loading group REPAIRED_HYBRID_TO_PANEL
IService::UPLOADDB                         INFO                Working on test code WIRE_BONDING
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
IService::log                              INFO                Logging status 1111 for component GPC1938_X_018_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC1938_X_018_A_H2/HybridBonding_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_WIRE_BONDING.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed GPC1938_X_018_A_H2.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//HybridBonding_v1//GPC1940_Y_008_B_H1.config
IService::next                             INFO                Scheduling HybridBonding_GPC1940_Y_008_B_H1.json file to run...
IService::next                             INFO                Writing results to /home/data/output//HybridBonding_v1//GPC1940_Y_008_B_H1/1/
MetadataValue
Date 2021-11-10T23:22:39.575+00:00
Operator Kirsten Affolder
Bonder hk2
Version v1
Bond numberTypeComment

Electrical results


Electrical summaries will be given below

ElectricalSvc::getListOfFilesFromDir       INFO                Scanning for test files GPC1938_X_018_A_H2/20USBHX2000778
ElectricalSvc::getListOfFilesFromDir       INFO                Found file GPC1938_X_018_A_H2_RC_4_6.txt
ElectricalSvc::getListOfFilesFromDir       INFO                Found file GPC1938_X_018_A_H2_RC_4_3.txt
ElectricalSvc::getListOfFilesFromDir       INFO                Found file GPC1938_X_018_A_H2_20220224.txt
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Scanning for test files GPC1938_X_018_A_H2/20USBHX2000778
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC1938_X_018_A_H2_20220224_4_16.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC1938_X_018_A_H2_20220224_4_6.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC1938_X_018_A_H2_20220224_4_3.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC1938_X_018_A_H2_20220224_4_2.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC1938_X_018_A_H2_20220224_4_1.json
ElectricalSvc::read                        INFO                Found 2 results files...
ElectricalSvc::read                        INFO                Analyzing test ResponseCurve for component GPC1938_X_018_A_H2
ElectricalSvc::read                        INFO                Detected 0 dead channels
ElectricalSvc::read                        INFO                Detected 0 stuck channels
ElectricalSvc::read                        INFO                Detected 0 low gain channels
ElectricalSvc::read                        INFO                Detected 0 high gain channels
ElectricalSvc::read                        INFO                Detected 0 low offset channels
ElectricalSvc::read                        INFO                Detected 0 high offset channels
ElectricalSvc::read                        INFO                Detected 2559 unbonded channels
ElectricalSvc::read                        INFO                Detected 0 part bonded channels
ElectricalSvc::read                        INFO                Detected 0 high noise channels
ElectricalSvc::read                        INFO                Detected 0 inefficient channels
ElectricalSvc::load_asic_db                INFO                Reading ASIC database file /home/data/upload//ElectricalHybrid//GPC1938_X_018_A_H2_GAIN_TEST.json
IService::UPLOADDB                         INFO                Working on test code PEDESTAL_TRIM_PPA
IService::UPLOADDB                         INFO                Working on test code STROBE_DELAY_PPA
IService::UPLOADDB                         INFO                Working on test code RESPONSE_CURVE_PPA
IService::log                              INFO                Logging status 1111 for component GPC1938_X_018_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC1938_X_018_A_H2/ElectricalHybrids-1/
IService::install                          INFO                File already installed: warningLog.txt...
IService::install                          INFO                File already installed: errorLog.txt...
IService::install                          INFO                File already installed: infoLog.txt...
IService::install                          INFO                File already installed: Gain_Channel1D.pdf...
IService::install                          INFO                File already installed: DB_READY_PEDESTAL_TRIM_PPA_GPC1938_X_018_A_H2_20220224_4_1.json...
IService::install                          INFO                File already installed: DB_READY_STROBE_DELAY_PPA_GPC1938_X_018_A_H2_20220224_4_2.json...
IService::install                          INFO                File already installed: DB_READY_RESPONSE_CURVE_PPA_GPC1938_X_018_A_H2_20220224_4_6.json...
IService::install                          INFO                File already installed: DB_READY_RESPONSE_CURVE_PPA_GPC1938_X_018_A_H2_20220224_4_3.json...
IService::install                          INFO                File already installed: GPC1938_X_018_A_H2.root...
IService::install                          INFO                File already installed: status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//ElectricalHybrid//GPC1940_Y_009_B_H4.config
IService::next                             INFO                Scheduling Panel-31-BondIDed file to run...
IService::next                             INFO                Writing results to /home/data/output//ElectricalHybrids//GPC1940_Y_009_B_H4/1/
DefectNumber of channels (%)
Dead 0
Stuck 0
Low gain 0
High gain 0
Low offset 0
High offset 0
Unbonded 2559
Part bonded 0
High noise 0
Inefficient 0

Figures of gain and noise, and correlations with wafer/hybrid level test results, if they exist.