Results for GPC2150_X_017_A_H5


Basic Information for GPC2150_X_017_A_H5


Brief description of DB components associated to this Hybrid Assembly and associated tests


Click for packaging sticker
MetadataValue
Alt Identifier UCSC-X-0120
Serial Number 20USBHX2001162
Assembled False
Completed True
Current Location SCIPP, UC Santa Cruz
Manufacture ID GPC2150_X_017_A_H5
User SCIPP Module Institute Account

Children

ComponentTypeTimestamp
HCC Chip HCCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
ABC Chip ABCSTAR 2022-12-09T20:12:16.592Z
STAR Hybrid Flex X 2022-12-09T20:12:19.710Z
HCC Chip HCCSTARV1 2022-12-18T22:37:22.929Z
ABC Chip ABCSTARV1 2022-12-18T22:37:23.483Z
ABC Chip ABCSTARV1 2022-12-18T22:37:24.045Z
ABC Chip ABCSTARV1 2022-12-18T22:37:24.584Z
ABC Chip ABCSTARV1 2022-12-18T22:37:25.083Z
ABC Chip ABCSTARV1 2022-12-18T22:37:25.643Z
ABC Chip ABCSTARV1 2022-12-18T22:37:26.168Z
ABC Chip ABCSTARV1 2022-12-18T22:37:26.692Z
ABC Chip ABCSTARV1 2022-12-18T22:37:27.218Z
ABC Chip ABCSTARV1 2022-12-18T22:37:27.720Z
ABC Chip ABCSTARV1 2022-12-18T22:37:28.352Z

Parents

ComponentTypeTimestamp
STAR Hybrid Test Panel
Module 2023-03-06T19:48:56.039Z

NameCurrent StageTimestamp

Test NamePassed?Date
Glue weight (ASICs on hybrid) 2022-12-09T20:25:26.863Z
ASIC Metrology 2022-12-10T00:35:19.086Z
Visual Inspection 2022-12-14T17:29:25.001Z
Wire bonding 2022-12-14T17:29:25.001Z
Noise Occupancy PPA
Strobe Delay PPA 2022-12-19T21:47:52.000Z
Response Curve PPA 2022-12-19T21:48:51.000Z
Pedestal Trim PPA 2022-12-19T21:47:44.000Z

Glue weight for GPC2150_X_017_A_H5


Shown below are results for the glue weight measurements

IService::UPLOADDB                         INFO                Working on test code ASIC_GLUE_WEIGHT
GlueWeightSvc::uploadDB                    INFO                Formatting database test for ASIC_GLUE_WEIGHT
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
IService::log                              INFO                Logging status 1111 for component GPC2150_X_017_A_H5
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2150_X_017_A_H5/GlueWeightHybridASIC_PPB_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed DB_COMMENTS.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_ASIC_GLUE_WEIGHT.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed GPC2150_X_017_A_H5.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//GlueWeightHybridASIC_PPB_v1//20USBHX2001163.config
IService::next                             INFO                Scheduling Data_20USBHX2001163.json file to run...
IService::next                             INFO                Writing results to /home/data/output//GlueWeightHybridASIC_PPB_v1//GPC2150_X_017_A_H6/1/

MetadataValue
Component type
Operator KA AND
Date 2022-12-09T20:25:26.863+00:00
Program ASIC-X-PPB.SRC
FieldWeight/Volume
Assummed density of glue 1.077
Total volume of ASIC glue [uL] 40.954
Volume per glue dot [nL] 787.582
Weight of glue under ASICs [mg] 44.100
Weights of bare hybrid [g] 2.074
Weights of stuffed hybrid [g] 2.118

Metrology results


Shown below are the results of the module metrology scans

MetrologySvc::analyzeHybridThickness       INFO                Total stack up 787.534750, chip thickness 300.000000 and glue 91.053250, giving hybrid 396.481500 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 783.801789, chip thickness 300.000000 and glue 89.244528, giving hybrid 394.557262 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 780.176927, chip thickness 300.000000 and glue 88.583417, giving hybrid 391.593510 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 777.412323, chip thickness 300.000000 and glue 86.736500, giving hybrid 390.675823 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 775.092795, chip thickness 300.000000 and glue 85.763556, giving hybrid 389.329239 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 775.032671, chip thickness 300.000000 and glue 82.652667, giving hybrid 392.380004 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 776.797356, chip thickness 300.000000 and glue 88.702167, giving hybrid 388.095190 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 773.293688, chip thickness 300.000000 and glue 89.774611, giving hybrid 383.519076 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 781.739732, chip thickness 300.000000 and glue 91.060833, giving hybrid 390.678899 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 783.152100, chip thickness 300.000000 and glue 99.410625, giving hybrid 383.741475 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 781.993000, chip thickness 300.000000 and glue 92.893625, giving hybrid 389.099375 thickness
IService::UPLOADDB                         INFO                Working on test code ASIC_METROLOGY
MetrologySvc::uploadDB                     INFO                Doing DB upload...
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyHybrid_PPB_X_v0//GPC2150_X_017_A_H5/1//Data__GPC2150_X_017_A_H5.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyHybrid_PPB_X_v0//GPC2150_X_017_A_H5/1//Results__GPC2150_X_017_A_H5.txt
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyHybrid_PPB_X_v0//GPC2150_X_017_A_H5/1//Data__GPC2150_X_017_A_H5.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyHybrid_PPB_X_v0//GPC2150_X_017_A_H5/1//Results__GPC2150_X_017_A_H5.txt
IService::log                              INFO                Logging status 1111 for component GPC2150_X_017_A_H5
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2150_X_017_A_H5/MetrologyHybrid_PPB_X_v0-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed Summary.pdf...
IService::install                          INFO                Installed TotalStackUp.pdf...
IService::install                          INFO                Installed HybridThickness.pdf...
IService::install                          INFO                Installed Positions.pdf...
IService::install                          INFO                Installed Tilt.pdf...
IService::install                          INFO                Installed Data__GPC2150_X_017_A_H5.txt...
IService::install                          INFO                Installed Results__GPC2150_X_017_A_H5.txt...
IService::install                          INFO                Installed Results.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_ASIC_METROLOGY.json...
IService::install                          INFO                Installed GPC2150_X_017_A_H5.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//MetrologyHybrid_PPB_X_v0//GPC2150_X_017_A_H6.config
IService::next                             INFO                Scheduling GPC2150-17-A-4-6-2149-8-B-0-2.txt file to run...
IService::next                             INFO                Writing results to /home/data/output//MetrologyHybrid_PPB_X_v0//GPC2150_X_017_A_H6/1/
MetaData::getLocalName                     WARNING             Could not find a local name for identifier 20USBHN0000631
Measurement::addPoint                      ERROR               Unknown type ?! Ignoring!
MetadataValue
Component type X
Operator Anthony Affolder
Date 2022-12-10T00:35:19.086+00:00
Program version PPB_cloneable_v0_DB102
Instrument SmartScope Flash 250

The thickness of the glue \(t_{glue}\) is calculated as: $$t_{glue} = \frac{1}{N_{c}}\sum_{c=1}^{N_{c}} z_{c} - \frac{1}{N_{b}}\sum_{b=1}^{N_{b}} z_{b} - t_{chip} $$ where \(t_{chip}\) is the thickness of the chip \(z_{i}\) is the z-height measurement, \(N_{i}\) is the number of measurements on a given surface, and \(c\) and \(b\) run over measurements taken on the chip and base, respectively. The assumed thickness of the chip is \(t_{chip} = 300\mu m\).

Uncertainties on the glue height measurement are estimated from the spread in z-height measurements on a given surface. On a given surface, the height uncertainty \(\sigma_{i}\) is calculated as: $$\sigma_{i} = \frac{\Delta z}{2\sqrt{N_{i}}}$$ where \(\Delta z = z_{max} - z_{min}\) and \(z_{max}\) and \(z_{min}\) are the maximum and minimum z-height measurements, respectively. In addition, a conservative \(\sigma_{chip} = 10\mu m\) uncertainty on the chip thickeness is assumed. The total uncertainty is calculated as: $$\sigma_{glue} = \sqrt{ \sigma_{c}^2 + \sigma_{b}^2 + \sigma_{chip} }$$ The glue thickness measurements for each ASIC are shown in the figure and table below.


The thickness total stack up is measured from the top of the each chip to the jig, and accounts for the combined thickness of hybrid flex, glue, and ABC/HCC chips in the case of hybrid.


The positions of the ASICs are measured on the 95x95 \(\mu m^2\) square pads on the opposite side of the sensor wire bonds. These measured positions are compared against the technical drawns to calculated the relative ASIC position. In some cases, the SmartScope incorrectly measures a different object than these square pads, and using properties (such as the perimeter), the position is corrected. The positions are shown in the figure below. If any corrections are applied, they are shown as transparent points.

The hybrid thickness is plotted below. It is calculated from the total stack up height minus the assembed chip thickness and measured glue thickness.

The tilt of a chip can be calculated in either the direction along the surface of a chip. We define the x-direction to be any points whose difference is much larger along x, and similarly for y. The tilt in both cases is defined to be: $$tilt = \frac{|\Delta z|}{\sqrt{\Delta x^{2} + \Delta y^{2}}} $$

Wirebonding for GPC2150_X_017_A_H5


Wirebonding summaries will be given below

WireBondSvc::run                           INFO                Loading group FAILED_ASIC_BACKEND
WireBondSvc::run                           INFO                Loading group FAILED_HYBRID_TO_PANEL
WireBondSvc::run                           INFO                Wirebond number(s) 761 with commment lifted off while moving to 762
WireBondSvc::run                           INFO                Loading group REPAIRED_ASIC_BACKEND
WireBondSvc::run                           INFO                Loading group REPAIRED_HYBRID_TO_PANEL
WireBondSvc::run                           INFO                Wirebond number(s) 761 with commment rebonded
IService::UPLOADDB                         INFO                Working on test code WIRE_BONDING
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
IService::log                              INFO                Logging status 1111 for component GPC2150_X_017_A_H5
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2150_X_017_A_H5/HybridBonding_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_WIRE_BONDING.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed GPC2150_X_017_A_H5.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//HybridBonding_v1//GPC2150_X_017_A_H6.config
IService::next                             INFO                Scheduling HybridBonding_GPC2150_X_017_A_H6.json file to run...
IService::next                             INFO                Writing results to /home/data/output//HybridBonding_v1//GPC2150_X_017_A_H6/1/
MetadataValue
Date 2022-12-14T17:30:19.072+00:00
Operator KA
Bonder hk2
Version v1
Bond numberTypeComment
761 FAILED_HYBRID_TO_PANEL lifted off while moving to 762
761 REPAIRED_HYBRID_TO_PANEL rebonded

Electrical results


Electrical summaries will be given below

ElectricalSvc::getListOfFilesFromDir       INFO                Scanning for test files GPC2150_X_017_A_H5/20USBHX2001162
ElectricalSvc::getListOfFilesFromDir       INFO                Found file GPC2150_X_017_A_H5_20221219.txt
ElectricalSvc::getListOfFilesFromDir       INFO                Found file GPC2150_X_017_A_H5_RC_7_4.txt
ElectricalSvc::getListOfFilesFromDir       INFO                Found file GPC2150_X_017_A_H5_RC_7_7.txt
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Scanning for test files GPC2150_X_017_A_H5/20USBHX2001162
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC2150_X_017_A_H5_20221219_7_2_PEDESTAL_TRIM_PPA.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC2150_X_017_A_H5_20221219_7_3_STROBE_DELAY_PPA.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC2150_X_017_A_H5_20221219_7_4_RESPONSE_CURVE_PPA.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC2150_X_017_A_H5_20221219_7_7_RESPONSE_CURVE_PPA.json
ElectricalSvc::getListOfJsonFilesFromDir   INFO                Loading json fle GPC2150_X_017_A_H5_20221219_7_17_NO_PPA.json
ElectricalSvc::read                        INFO                Found 2 results files...
ElectricalSvc::read                        INFO                Analyzing test ResponseCurve for component GPC2150_X_017_A_H5
ElectricalSvc::read                        INFO                Detected 0 dead channels
ElectricalSvc::read                        INFO                Detected 0 stuck channels
ElectricalSvc::read                        INFO                Detected 0 low gain channels
ElectricalSvc::read                        INFO                Detected 0 high gain channels
ElectricalSvc::read                        INFO                Detected 0 low offset channels
ElectricalSvc::read                        INFO                Detected 0 high offset channels
ElectricalSvc::read                        INFO                Detected 2557 unbonded channels
ElectricalSvc::read                        INFO                Detected 0 part bonded channels
ElectricalSvc::read                        INFO                Detected 0 high noise channels
ElectricalSvc::read                        INFO                Detected 0 inefficient channels
ElectricalSvc::load_asic_db                INFO                Reading ASIC database file /home/data/upload//ElectricalHybrids_PPB_v1//GPC2150_X_017_A_H5_GAIN_TEST.json
IService::UPLOADDB                         INFO                Working on test code PEDESTAL_TRIM_PPA
IService::UPLOADDB                         INFO                Working on test code STROBE_DELAY_PPA
IService::UPLOADDB                         INFO                Working on test code RESPONSE_CURVE_PPA
IService::UPLOADDB                         INFO                Working on test code NO_PPA
IService::log                              INFO                Logging status 1111 for component GPC2150_X_017_A_H5
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2150_X_017_A_H5/ElectricalHybrids_PPB_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed Gain_Channel1D.pdf...
IService::install                          INFO                Installed DB_READY_PEDESTAL_TRIM_PPA_GPC2150_X_017_A_H5_20221219_7_2_PEDESTAL_TRIM_PPA.json...
IService::install                          INFO                Installed DB_READY_STROBE_DELAY_PPA_GPC2150_X_017_A_H5_20221219_7_3_STROBE_DELAY_PPA.json...
IService::install                          INFO                Installed DB_READY_RESPONSE_CURVE_PPA_GPC2150_X_017_A_H5_20221219_7_4_RESPONSE_CURVE_PPA.json...
IService::install                          INFO                Installed DB_READY_RESPONSE_CURVE_PPA_GPC2150_X_017_A_H5_20221219_7_7_RESPONSE_CURVE_PPA.json...
IService::install                          INFO                Installed DB_READY_NO_PPA_GPC2150_X_017_A_H5_20221219_7_17_NO_PPA.json...
IService::install                          INFO                Installed GPC2150_X_017_A_H5.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//ElectricalHybrids_PPB_v1//GPC2150_X_017_A_H6.config
IService::next                             INFO                Scheduling Panel030_Dec19 file to run...
IService::next                             INFO                Writing results to /home/data/output//ElectricalHybrids_PPB_v1//GPC2150_X_017_A_H6/1/
MetaData::getLocalName                     WARNING             Could not find a local name for identifier 20USBHN0000631
DefectNumber of channels (%)
Dead 0
Stuck 0
Low gain 0
High gain 0
Low offset 0
High offset 0
Unbonded 2557
Part bonded 0
High noise 0
Inefficient 0

Figures of gain and noise, and correlations with wafer/hybrid level test results, if they exist.