The information on this page was generated on Sun Apr 5 11:21:25 2026 (probably UTC). You may click the button below to manually regenerate the content.
Basic Information for GPC2225_X_015_B_H0
Brief description of DB components associated to this Hybrid Assembly and associated tests
Shown below are results for the glue weight measurements
IService::UPLOADDB INFO Working on test code ASIC_GLUE_WEIGHT
GlueWeightSvc::uploadDB INFO Formatting database test for ASIC_GLUE_WEIGHT
IService::UPLOADDB INFO Working on test code VISUAL_INSPECTION
IService::log INFO Logging status 0111 for component GPC2225_X_015_B_H0
IService::install INFO Installing output directory to /home/data/modules/GPC2225_X_015_B_H0/GlueWeightHybridASIC_PROD_v2-1/
IService::install INFO Installed warningLog.txt...
IService::install INFO Installed errorLog.txt...
IService::install INFO Installed infoLog.txt...
IService::install INFO Installed 20USBHX2002230_IMG_0109.jpeg...
IService::install INFO Installed DB_COMMENTS.json...
IService::install INFO Installed DB_READY_HYBRID_ASSEMBLY_ASIC_GLUE_WEIGHT.json...
IService::install INFO Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION.json...
IService::install INFO Installed GPC2225_X_015_B_H0.root...
IService::install INFO Installed status.json...
TopControl::monitor INFO Service with algoirthm name GlueWeightModule_LS_PPA ignored through users request!
TopControl::monitor INFO Service with algoirthm name GlueWeightModule_SS_PPA ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridBonding_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name ModuleBonding_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridPullTest ignored through users request!
TopControl::monitor INFO Service with algoirthm name PullTestCicorelCard ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridFlexArray_PullTest_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name Powerboards_PullTest_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name ABC_PullTest_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name ABC_Wafer_PullTest_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name HCC_PullTest_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name SensorBondingQC_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name HVTabSheet_SheerTest_Before_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name HVTabSheet_SheerTest_After_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name PhotoUpload ignored through users request!
TopControl::monitor INFO Service with algoirthm name PictureModule ignored through users request!
TopControl::monitor INFO Service with algoirthm name GeneralPictures ignored through users request!
TopControl::monitor INFO Service with algoirthm name PreHVTabIV ignored through users request!
TopControl::monitor INFO Service with algoirthm name SensorIV_PreTabIV_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name Module_postHVTabIV_LS_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name SensorIV_PostTabIV_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name ElectricalIV ignored through users request!
TopControl::monitor INFO Service with algoirthm name SensorIV_Module_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name SensorIV_AMACIV_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name SixWayJig_Y_slots ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridMetrology ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridPanelMetrology ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridPanelMetrology_v7 ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridMetrology_v8 ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridMetrology_Y_v8 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_v10 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_v10-beta ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin_interposer ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_PROD_X_v2 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_PROD_X_v2_interposer ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick_interposer ignored through users request!
TopControl::monitor INFO Service with algoirthm name HybridToSensorMetrology_v2a ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_v8 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_v8a ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PPA_SS_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PPB_SS_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PPB_SS_wuerth_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PPB_LS_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PROD_LS_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PROD_SS_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PROD_LS_v1_thin ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PROD_LS_v1_interposer ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PROD_LS_v1_thin_interposer ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PROD_SS_v1_thin ignored through users request!
TopControl::monitor INFO Service with algoirthm name MetrologyModules_PROD_SS_v1_thin_interposer ignored through users request!
TopControl::monitor INFO Service with algoirthm name ModuleLS_Bow_PPA_v0 ignored through users request!
TopControl::monitor INFO Service with algoirthm name ModuleSS_Bow_PPA_v0 ignored through users request!
TopControl::monitor INFO Service with algoirthm name ElectricalHybrids ignored through users request!
TopControl::monitor INFO Service with algoirthm name ElectricalHybrids_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name BurnInHybrids_PPB_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name BurnInHybrids_PROD_v2 ignored through users request!
TopControl::monitor INFO Service with algoirthm name ElectricalModules_PPA_v1 ignored through users request!
TopControl::monitor INFO Service with algoirthm name TCModules_PROD_v0 ignored through users request!
Metadata
Value
Component type
Operator
QFP, LM
Date
2025-03-06T23:48:45.868+00:00
Program
ASIC_X_PPB.SRC
Field
Weight/Volume
Assummed density of glue
1.077
Total volume of ASIC glue [uL]
40.304
Volume per glue dot [nL]
775.081
Weight of glue under ASICs [mg]
43.400
Weights of bare hybrid [g]
2.898
Weights of stuffed hybrid [g]
2.942
Metrology results
Shown below are the results of the module metrology scans
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_0] Height in z : 230.789250 +/- 5.308000 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_0] Height in z : -179.748500 +/- 13.299264 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 410.537750
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 110.537750 +/- 15.167244 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 411.537750 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_1] Height in z : 255.252000 +/- 5.945500 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_1] Height in z : -148.422667 +/- 8.129333 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 403.674667
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 103.674667 +/- 11.244333 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 404.674667 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_2] Height in z : 276.267750 +/- 5.181000 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_2] Height in z : -121.168333 +/- 6.264167 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 397.436083
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 97.436083 +/- 9.543718 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 398.436083 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_3] Height in z : 296.860500 +/- 4.704500 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_3] Height in z : -100.330889 +/- 4.945333 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 397.191389
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 97.191389 +/- 8.461007 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 398.191389 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_4] Height in z : 321.254500 +/- 4.613250 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_4] Height in z : -73.789444 +/- 5.654667 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 395.043944
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 95.043944 +/- 8.846317 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 396.043944 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_5] Height in z : 342.415000 +/- 5.463500 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_5] Height in z : -52.318444 +/- 6.273333 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 394.733444
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 94.733444 +/- 9.705903 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 395.733444 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_6] Height in z : 368.926500 +/- 5.705250 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_6] Height in z : -23.037556 +/- 6.471167 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 391.964056
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 91.964056 +/- 9.971252 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 392.964056 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_7] Height in z : 394.726750 +/- 6.539500 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_7] Height in z : 5.086222 +/- 6.509667 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 389.640528
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 89.640528 +/- 10.494800 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 390.640528 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_8] Height in z : 423.326000 +/- 5.937250 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_8] Height in z : 29.360556 +/- 9.708000 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 393.965444
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 93.965444 +/- 12.429650 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 394.965444 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_9] Height in z : 452.424500 +/- 6.174750 um
MetrologySvc::analyzeGlueHeights INFO ----> [ABC_X_9] Height in z : 52.128875 +/- 9.004651 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 400.295625
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 100.295625 +/- 12.008800 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 401.295625 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeGlueHeights INFO Printing associated surfaces...
MetrologySvc::analyzeGlueHeights INFO ----> [HCC_X_0] Height in z : 431.167750 +/- 4.750750 um
MetrologySvc::analyzeGlueHeights INFO ----> [HCC_X_0] Height in z : 34.460375 +/- 8.456467 um
MetrologySvc::analyzeGlueHeights INFO ----> Difference in heights : 396.707375
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 96.707375 +/- 10.912445 um
MetrologySvc::analyzeGlueHeights INFO ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 397.707375 um
MetrologySvc::analyzeGlueHeights INFO ---------------------------------------------------------------------------------
MetrologySvc::analyzeHybridThickness INFO Total stack up 1000.780750, chip thickness 300.000000 and glue 110.537750, giving hybrid 590.243000 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 1000.471430, chip thickness 300.000000 and glue 103.674667, giving hybrid 596.796763 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 997.177926, chip thickness 300.000000 and glue 97.436083, giving hybrid 599.741842 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 993.143808, chip thickness 300.000000 and glue 97.191389, giving hybrid 595.952419 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 993.517640, chip thickness 300.000000 and glue 95.043944, giving hybrid 598.473695 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 990.301786, chip thickness 300.000000 and glue 94.733444, giving hybrid 595.568341 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 990.933501, chip thickness 300.000000 and glue 91.964056, giving hybrid 598.969446 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 989.275517, chip thickness 300.000000 and glue 89.640528, giving hybrid 599.634989 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 990.206773, chip thickness 300.000000 and glue 93.965444, giving hybrid 596.241329 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 992.867214, chip thickness 300.000000 and glue 100.295625, giving hybrid 592.571589 thickness
MetrologySvc::analyzeHybridThickness INFO Total stack up 987.242750, chip thickness 300.000000 and glue 96.707375, giving hybrid 590.535375 thickness
IService::UPLOADDB INFO Working on test code ASIC_METROLOGY
MetrologySvc::uploadDB INFO Doing DB upload...
MetrologySvc::writeOutput INFO Writing common output file /home/data/output//MetrologyHybrid_PROD_X_v2_thick_interposer//GPC2225_X_015_B_H0/1//Data__GPC2225_X_015_B_H0.txt
MetrologySvc::writeOutput INFO Writing results output file /home/data/output//MetrologyHybrid_PROD_X_v2_thick_interposer//GPC2225_X_015_B_H0/1//Results__GPC2225_X_015_B_H0.txt
MetrologySvc::writeOutput INFO Writing common output file /home/data/output//MetrologyHybrid_PROD_X_v2_thick_interposer//GPC2225_X_015_B_H0/1//Data__GPC2225_X_015_B_H0.txt
MetrologySvc::writeOutput INFO Writing results output file /home/data/output//MetrologyHybrid_PROD_X_v2_thick_interposer//GPC2225_X_015_B_H0/1//Results__GPC2225_X_015_B_H0.txt
IService::log INFO Logging status 0111 for component GPC2225_X_015_B_H0
IService::install INFO Installing output directory to /home/data/modules/GPC2225_X_015_B_H0/MetrologyHybrid_PROD_X_v2_thick_interposer-1/
IService::install INFO Installed warningLog.txt...
IService::install INFO Installed errorLog.txt...
IService::install INFO Installed infoLog.txt...
IService::install INFO Installed Summary.pdf...
IService::install INFO Installed TotalStackUp.pdf...
IService::install INFO Installed HybridThickness.pdf...
IService::install INFO Installed Positions.pdf...
IService::install INFO Installed Tilt.pdf...
IService::install INFO Installed Data__GPC2225_X_015_B_H0.txt...
IService::install INFO Installed Results__GPC2225_X_015_B_H0.txt...
IService::install INFO Installed Results.json...
IService::install INFO Installed DB_READY_HYBRID_ASSEMBLY_ASIC_METROLOGY.json...
IService::install INFO Installed GPC2225_X_015_B_H0.root...
IService::install INFO Installed status.json...
MetaData::readField INFO Reading file /home/data/upload//MetrologyHybrid_PROD_X_v2_thick_interposer//GPC2225_X_015_B_H1-1.config
IService::next INFO Scheduling GPC2225-X-015-B0-B1-B2-B3-B6_2225-X-010-B1.txt file to run...
IService::next INFO Writing results to /home/data/output//MetrologyHybrid_PROD_X_v2_thick_interposer//GPC2225_X_015_B_H1/1/
MetaData::getLocalName WARNING Could not find a local name for identifier 20USBHN0001447
The thickness of the glue \(t_{glue}\) is calculated as:
$$t_{glue} = \frac{1}{N_{c}}\sum_{c=1}^{N_{c}} z_{c} - \frac{1}{N_{b}}\sum_{b=1}^{N_{b}} z_{b} - t_{chip} $$
where \(t_{chip}\) is the thickness of the chip
\(z_{i}\) is the z-height measurement,
\(N_{i}\) is the number of measurements on a given surface,
and \(c\) and \(b\) run over measurements taken on the chip and base, respectively.
The assumed thickness of the chip is \(t_{chip} = 300\mu m\).
Uncertainties on the glue height measurement are estimated from the spread in z-height measurements on a given surface.
On a given surface, the height uncertainty \(\sigma_{i}\) is calculated as:
$$\sigma_{i} = \frac{\Delta z}{2\sqrt{N_{i}}}$$
where \(\Delta z = z_{max} - z_{min}\) and \(z_{max}\) and \(z_{min}\) are the maximum and minimum z-height measurements, respectively.
In addition, a conservative \(\sigma_{chip} = 10\mu m\) uncertainty on the chip thickeness is assumed.
The total uncertainty is calculated as:
$$\sigma_{glue} = \sqrt{ \sigma_{c}^2 + \sigma_{b}^2 + \sigma_{chip} }$$
The glue thickness measurements for each ASIC are shown in the figure and table below.
The thickness total stack up is measured from the top of the each chip to the jig,
and accounts for the combined thickness of hybrid flex, glue, and ABC/HCC chips in the case of hybrid.
The positions of the ASICs are measured on the 95x95 \(\mu m^2\) square pads on the opposite side of the sensor wire bonds.
These measured positions are compared against the technical drawns to calculated the relative ASIC position.
In some cases, the SmartScope incorrectly measures a different object than these square pads,
and using properties (such as the perimeter), the position is corrected.
The positions are shown in the figure below. If any corrections are applied, they are shown as transparent points.
The hybrid thickness is plotted below. It is calculated from the total stack up height minus the assembed chip thickness and measured glue thickness.
The tilt of a chip can be calculated in either the direction along the surface of a chip. We define
the x-direction to be any points whose difference is much larger along x, and similarly for y. The tilt in both cases is defined to be:
$$tilt = \frac{|\Delta z|}{\sqrt{\Delta x^{2} + \Delta y^{2}}} $$
Wirebonding for GPC2225_X_015_B_H0
Wirebonding summaries will be given below
WireBondSvc::run INFO Loading group FAILED_ASIC_BACKEND
WireBondSvc::run INFO Loading group FAILED_HYBRID_TO_PANEL
WireBondSvc::run INFO Loading group REPAIRED_ASIC_BACKEND
WireBondSvc::run INFO Loading group REPAIRED_HYBRID_TO_PANEL
IService::UPLOADDB INFO Working on test code WIRE_BONDING
IService::UPLOADDB INFO Working on test code VISUAL_INSPECTION
IService::log INFO Logging status 0111 for component GPC2225_X_015_B_H0
IService::install INFO Installing output directory to /home/data/modules/GPC2225_X_015_B_H0/HybridBonding_v1-1/
IService::install INFO Installed warningLog.txt...
IService::install INFO Installed errorLog.txt...
IService::install INFO Installed infoLog.txt...
IService::install INFO Installed DB_READY_HYBRID_ASSEMBLY_WIRE_BONDING.json...
IService::install INFO Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION.json...
IService::install INFO Installed GPC2225_X_015_B_H0.root...
IService::install INFO Installed status.json...
MetaData::readField INFO Reading file /home/data/upload//HybridBonding_v1//GPC2225_X_015_B_H1-1.config
IService::next INFO Scheduling HybridBonding_GPC2225_X_015_B_H1-1.json file to run...
IService::next INFO Writing results to /home/data/output//HybridBonding_v1//GPC2225_X_015_B_H1/1/