Results for GPC2349_X_010_A_H2


The information on this page was generated on Wed Apr 1 11:57:36 2026 (probably UTC).
You may click the button below to manually regenerate the content.

Basic Information for GPC2349_X_010_A_H2


Brief description of DB components associated to this Hybrid Assembly and associated tests


Click for packaging sticker
MetadataValue
Alt Identifier UCSC-X-0813
Serial Number 20USBHX2004667
Batch PRODUCTION_UCSC
Assembled True
Completed False
Current Location SCIPP, UC Santa Cruz
Manufacture ID GPC2349_X_010_A_H2
User SCIPP Module Institute Account

Children

ComponentTypeTimestamp
STAR Hybrid Flex X 2026-01-21T23:49:07.299Z
HCC Chip HCCSTARV1 2026-01-27T23:54:56.863Z
ABC Chip ABCSTARV1 2026-01-27T23:54:58.641Z
ABC Chip ABCSTARV1 2026-01-27T23:55:00.252Z
ABC Chip ABCSTARV1 2026-01-27T23:55:02.077Z
ABC Chip ABCSTARV1 2026-01-27T23:55:03.711Z
ABC Chip ABCSTARV1 2026-01-27T23:55:05.305Z
ABC Chip ABCSTARV1 2026-01-27T23:55:06.952Z
ABC Chip ABCSTARV1 2026-01-27T23:55:08.548Z
ABC Chip ABCSTARV1 2026-01-27T23:55:10.114Z
ABC Chip ABCSTARV1 2026-01-27T23:55:11.925Z
ABC Chip ABCSTARV1 2026-01-27T23:55:13.556Z
Hybrid Interposer 2026-01-21T23:49:04.664Z

Parents

ComponentTypeTimestamp
STAR Hybrid Test Panel
Module 2026-02-03T19:45:24.790Z

NameCurrent StageTimestamp

Test NamePassed?Date
Glue weight (ASICs on hybrid) 2026-01-22T00:06:47.678Z
Visual Inspection 2026-01-23T22:34:06.230Z
ASIC Metrology 2026-01-22T21:06:03.971Z
Wire bonding 2026-01-23T22:34:06.230Z
Pedestal Trim TC 2026-02-12T13:55:52.000Z
Strobe Delay TC 2026-02-12T13:56:10.000Z
Response Curve TC 2026-02-12T14:04:31.000Z
Noise Occupancy TC 2026-02-12T14:07:08.000Z
Open Channel Search TC 2026-02-12T14:10:26.000Z

Glue weight for GPC2349_X_010_A_H2


Shown below are results for the glue weight measurements

IService::UPLOADDB                         INFO                Working on test code ASIC_GLUE_WEIGHT
GlueWeightSvc::uploadDB                    INFO                Formatting database test for ASIC_GLUE_WEIGHT
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
IService::log                              INFO                Logging status 0111 for component GPC2349_X_010_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2349_X_010_A_H2/GlueWeightHybridASIC_PROD_v2-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed 20USBHX2004667_image.jpg...
IService::install                          INFO                Installed DB_COMMENTS.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_ASIC_GLUE_WEIGHT.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed GPC2349_X_010_A_H2.root...
IService::install                          INFO                Installed status.json...
TopControl::monitor                        INFO                Service with algoirthm name GlueWeightModule_LS_PPA ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name GlueWeightModule_SS_PPA ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridBonding_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleBonding_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPullTest ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PullTestCicorelCard ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridFlexArray_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name Powerboards_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ABC_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ABC_Wafer_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HCC_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorBondingQC_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HVTabSheet_SheerTest_Before_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HVTabSheet_SheerTest_After_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PhotoUpload ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PictureModule ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name GeneralPictures ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PreHVTabIV ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_PreTabIV_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name Module_postHVTabIV_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_PostTabIV_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalIV ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_Module_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_AMACIV_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SixWayJig_Y_slots ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPanelMetrology ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPanelMetrology_v7 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology_Y_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_v10 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_v10-beta ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridToSensorMetrology_v2a ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_v8a ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPA_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_SS_wuerth_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleLS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleSS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PROD_v2 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalModules_PPA_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name TCModules_PROD_v0 ignored through users request!

MetadataValue
Component type
Operator AS
Date 2026-01-22T00:06:47.678+00:00
Program ASIC_X_PPB.SRC
FieldWeight/Volume
Assummed density of glue 1.077
Total volume of ASIC glue [uL] 41.233
Volume per glue dot [nL] 792.940
Weight of glue under ASICs [mg] 44.400
Weights of bare hybrid [g] 2.537
Weights of stuffed hybrid [g] 2.581

Metrology results


Shown below are the results of the module metrology scans

MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_0] Height in z :                               440.257000 +/- 7.342250 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_0] Height in z :                               55.952000 +/- 7.663623 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             384.305000
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 84.305000 +/- 11.731997 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 385.305000 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_1] Height in z :                               441.391750 +/- 0.599750 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_1] Height in z :                               58.655556 +/- 8.905167 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             382.736194
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 82.736194 +/- 10.230430 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 383.736194 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_2] Height in z :                               438.006750 +/- 1.489250 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_2] Height in z :                               53.786778 +/- 7.938333 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             384.219972
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 84.219972 +/- 9.499211 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 385.219972 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_3] Height in z :                               434.247750 +/- 1.575500 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_3] Height in z :                               47.536889 +/- 9.527333 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             386.710861
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 86.710861 +/- 10.874386 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 387.710861 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_4] Height in z :                               432.363750 +/- 1.934000 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_4] Height in z :                               59.631667 +/- 8.459333 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             372.732083
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 72.732083 +/- 10.015023 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 373.732083 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_5] Height in z :                               429.916500 +/- 1.426250 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_5] Height in z :                               59.427444 +/- 7.342833 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             370.489056
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 70.489056 +/- 8.997299 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 371.489056 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_6] Height in z :                               429.450750 +/- 1.421250 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_6] Height in z :                               55.083333 +/- 7.429000 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             374.367417
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 74.367417 +/- 9.066973 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 375.367417 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_7] Height in z :                               428.847000 +/- 1.574500 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_7] Height in z :                               49.610778 +/- 7.236000 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             379.236222
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 79.236222 +/- 8.935253 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 380.236222 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_8] Height in z :                               429.793250 +/- 2.130500 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_8] Height in z :                               49.593000 +/- 7.255833 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             380.200250
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 80.200250 +/- 9.065658 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 381.200250 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_9] Height in z :                               432.863250 +/- 1.389500 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_9] Height in z :                               52.500250 +/- 9.642815 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             380.363000
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 80.363000 +/- 10.950552 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 381.363000 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [HCC_X_0] Height in z :                               434.132000 +/- 0.890500 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [HCC_X_0] Height in z :                               48.728750 +/- 4.676451 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             385.403250
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 300.000000 +/- 5.000000 extra material : 85.403250 +/- 6.903780 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 386.403250 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 944.138000, chip thickness 300.000000 and glue 84.305000, giving hybrid 559.833000 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 939.667202, chip thickness 300.000000 and glue 82.736194, giving hybrid 556.931008 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 937.778427, chip thickness 300.000000 and glue 84.219972, giving hybrid 553.558455 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 935.204522, chip thickness 300.000000 and glue 86.710861, giving hybrid 548.493661 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 935.105454, chip thickness 300.000000 and glue 72.732083, giving hybrid 562.373371 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 934.100713, chip thickness 300.000000 and glue 70.489056, giving hybrid 563.611657 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 933.568914, chip thickness 300.000000 and glue 74.367417, giving hybrid 559.201498 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 931.282195, chip thickness 300.000000 and glue 79.236222, giving hybrid 552.045973 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 930.344405, chip thickness 300.000000 and glue 80.200250, giving hybrid 550.144155 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 932.799154, chip thickness 300.000000 and glue 80.363000, giving hybrid 552.436154 thickness
MetrologySvc::analyzeHybridThickness       INFO                Total stack up 934.838750, chip thickness 300.000000 and glue 85.403250, giving hybrid 549.435500 thickness
IService::UPLOADDB                         INFO                Working on test code ASIC_METROLOGY
MetrologySvc::uploadDB                     INFO                Doing DB upload...
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyHybrid_PROD_X_v2_interposer//GPC2349_X_010_A_H2/1//Data__GPC2349_X_010_A_H2.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyHybrid_PROD_X_v2_interposer//GPC2349_X_010_A_H2/1//Results__GPC2349_X_010_A_H2.txt
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyHybrid_PROD_X_v2_interposer//GPC2349_X_010_A_H2/1//Data__GPC2349_X_010_A_H2.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyHybrid_PROD_X_v2_interposer//GPC2349_X_010_A_H2/1//Results__GPC2349_X_010_A_H2.txt
IService::log                              INFO                Logging status 0111 for component GPC2349_X_010_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2349_X_010_A_H2/MetrologyHybrid_PROD_X_v2_interposer-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed Summary.pdf...
IService::install                          INFO                Installed TotalStackUp.pdf...
IService::install                          INFO                Installed HybridThickness.pdf...
IService::install                          INFO                Installed Positions.pdf...
IService::install                          INFO                Installed Tilt.pdf...
IService::install                          INFO                Installed Data__GPC2349_X_010_A_H2.txt...
IService::install                          INFO                Installed Results__GPC2349_X_010_A_H2.txt...
IService::install                          INFO                Installed Results.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_ASIC_METROLOGY.json...
IService::install                          INFO                Installed GPC2349_X_010_A_H2.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//MetrologyHybrid_PROD_X_v2_interposer//GPC2349_X_010_A_H4-1.config
IService::next                             INFO                Scheduling GPC2349-008-B4-5-6_GPC2349-010-A0-2-4.txt file to run...
IService::next                             INFO                Writing results to /home/data/output//MetrologyHybrid_PROD_X_v2_interposer//GPC2349_X_010_A_H4/1/
MetaData::getLocalName                     WARNING             Could not find a local name for identifier 20USBHN0006881
MetadataValue
Component type X
Operator LM
Date 2026-01-22T21:06:03.971+00:00
Program version MetrologyHybrid_PROD_X_v2_interposer/PROD_clonable_v2_new-jig_DB102_6hybrids_thin_interposer
Instrument SmartScope Flash 250

The thickness of the glue \(t_{glue}\) is calculated as: $$t_{glue} = \frac{1}{N_{c}}\sum_{c=1}^{N_{c}} z_{c} - \frac{1}{N_{b}}\sum_{b=1}^{N_{b}} z_{b} - t_{chip} $$ where \(t_{chip}\) is the thickness of the chip \(z_{i}\) is the z-height measurement, \(N_{i}\) is the number of measurements on a given surface, and \(c\) and \(b\) run over measurements taken on the chip and base, respectively. The assumed thickness of the chip is \(t_{chip} = 300\mu m\).

Uncertainties on the glue height measurement are estimated from the spread in z-height measurements on a given surface. On a given surface, the height uncertainty \(\sigma_{i}\) is calculated as: $$\sigma_{i} = \frac{\Delta z}{2\sqrt{N_{i}}}$$ where \(\Delta z = z_{max} - z_{min}\) and \(z_{max}\) and \(z_{min}\) are the maximum and minimum z-height measurements, respectively. In addition, a conservative \(\sigma_{chip} = 10\mu m\) uncertainty on the chip thickeness is assumed. The total uncertainty is calculated as: $$\sigma_{glue} = \sqrt{ \sigma_{c}^2 + \sigma_{b}^2 + \sigma_{chip} }$$ The glue thickness measurements for each ASIC are shown in the figure and table below.


The thickness total stack up is measured from the top of the each chip to the jig, and accounts for the combined thickness of hybrid flex, glue, and ABC/HCC chips in the case of hybrid.


The positions of the ASICs are measured on the 95x95 \(\mu m^2\) square pads on the opposite side of the sensor wire bonds. These measured positions are compared against the technical drawns to calculated the relative ASIC position. In some cases, the SmartScope incorrectly measures a different object than these square pads, and using properties (such as the perimeter), the position is corrected. The positions are shown in the figure below. If any corrections are applied, they are shown as transparent points.

The hybrid thickness is plotted below. It is calculated from the total stack up height minus the assembed chip thickness and measured glue thickness.

The tilt of a chip can be calculated in either the direction along the surface of a chip. We define the x-direction to be any points whose difference is much larger along x, and similarly for y. The tilt in both cases is defined to be: $$tilt = \frac{|\Delta z|}{\sqrt{\Delta x^{2} + \Delta y^{2}}} $$

Wirebonding for GPC2349_X_010_A_H2


Wirebonding summaries will be given below

WireBondSvc::run                           INFO                Loading group FAILED_ASIC_BACKEND
WireBondSvc::run                           INFO                Loading group FAILED_HYBRID_TO_PANEL
WireBondSvc::run                           INFO                Loading group REPAIRED_ASIC_BACKEND
WireBondSvc::run                           INFO                Loading group REPAIRED_HYBRID_TO_PANEL
IService::UPLOADDB                         INFO                Working on test code WIRE_BONDING
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
IService::log                              INFO                Logging status 0111 for component GPC2349_X_010_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2349_X_010_A_H2/HybridBonding_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_WIRE_BONDING.json...
IService::install                          INFO                Installed DB_READY_HYBRID_ASSEMBLY_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed GPC2349_X_010_A_H2.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//HybridBonding_v1//GPC2349_X_010_A_H3-1.config
IService::next                             INFO                Scheduling HybridBonding_GPC2349_X_010_A_H3-1.json file to run...
IService::next                             INFO                Writing results to /home/data/output//HybridBonding_v1//GPC2349_X_010_A_H3/1/
MetadataValue
Date 2026-01-23T22:34:06.230+00:00
Operator LM
Bonder hk1
Version v1
Bond numberTypeComment

Electrical results


Electrical summaries will be given below

ElectricalSvc::read                        INFO                Found 0 results files...
ElectricalSvc::load_asic_db                INFO                Reading ASIC database file /home/data/upload//BurnInHybrids_PROD_v2//GPC2349_X_010_A_H2_GAIN_TEST.json
ElectricalSvc::PassFail                    INFO                Reading PassFail status directly from uploaded files
IService::UPLOADDB                         INFO                Working on test code PEDESTAL_TRIM_BURNIN
IService::UPLOADDB                         INFO                Working on test code STROBE_DELAY_BURNIN
IService::UPLOADDB                         INFO                Working on test code RESPONSE_CURVE_BURNIN
IService::UPLOADDB                         INFO                Working on test code NO_BURNIN
IService::log                              INFO                Logging status 0111 for component GPC2349_X_010_A_H2
IService::install                          INFO                Installing output directory to /home/data/modules/GPC2349_X_010_A_H2/BurnInHybrids_PROD_v2-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed GPC2349_X_010_A_H2.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//BurnInHybrids_PROD_v2//GPC2349_X_010_A_H3-1.config
IService::next                             INFO                Scheduling burnin_results_54_jan27 file to run...
IService::next                             INFO                Writing results to /home/data/output//BurnInHybrids_PROD_v2//GPC2349_X_010_A_H3/1/
ElectricalSvc::load_asic_db                WARNING             Could not find database ASIC GAIN_TEST results for GPC2349_X_010_A_H2
MetaData::getLocalName                     WARNING             Could not find a local name for identifier 20USBHN0006881
ElectricalSvc::getListOfFilesFromDir       ERROR               Could not open directory /home/data/upload//BurnInHybrids_PROD_v2//burnin_results_54_jan27/results/ to collect files. Returning no files...
ElectricalSvc::getListOfJsonFilesFromDir   ERROR               Could not open directory /home/data/upload//BurnInHybrids_PROD_v2//burnin_results_54_jan27/results/ to collect files. Returning no files...

Figures of gain and noise, and correlations with wafer/hybrid level test results, if they exist.