Results for SCIPP-PRO_LS-0298


The information on this page was generated on Sat Dec 27 08:50:28 2025 (probably UTC).
You may click the button below to manually regenerate the content.

Photos of GeneralPicture_Module_2025-11-18_20-14-20


SCIPP-PRO_LS-0298 after gluing

Photos of GeneralPicture_Module_2025-11-21_00-38-22


SCIPP-PRO_LS-0298 after wirebonding

Basic Information for SCIPP-PRO_LS-0298


Brief description of DB components associated to this Module and associated tests


Click for packaging sticker
MetadataValue
Alt Identifier UCSC-BARREL_LS_MODULE-0491
Serial Number 20USBML1236957
Batch PRODUCTION_UCSC
Assembled True
Completed False
Current Location SCIPP, UC Santa Cruz
Manufacture ID SCIPP-PRO_LS-0298
User SCIPP Module Institute Account

Children

Alt IdentifierLocal NameSerial NumberTypeTrashedCompletedTimestamp
20USBP05010661 B3 False True 2025-11-18T18:14:44.791Z
UCSC-X-0633 GPC2328_X_019_B_H1 20USBHX2003449 X False False 2025-11-14T22:16:47.824Z
VPA38892-W02331 20USBSL0002331 ATLAS18LS False True 2025-10-31T17:59:24.255Z

Parents

ComponentTypeTimestamp
Module Frame 2025-10-31T17:59:26.114Z

NameCurrent StageTimestamp

Test NamePassed?Date
Module IV with PS V1 2025-11-14T18:10:44.001Z
Visual Inspection 2025-11-21T00:26:39.134Z
Module glue weights 2025-11-18T19:19:30.671Z
Module Metrology 2025-11-18T18:17:48.773Z
Module Wire Bonding 2025-11-21T00:26:39.134Z
Module AMAC IV TC 2025-12-06T15:33:35.000Z
Module Thermal Cycling 2025-12-05T13:37:35.000Z

HVTab IV for SCIPP-PRO_LS-0298


Shown below are results for the HVTab IV

SensorIVSvc::read                          INFO                Using measurement with voltage 700.100000 and current 0.082950 for current stability calculation
SensorIVSvc::read                          INFO                Using measurement with voltage 700.120000 and current 0.081450 for current stability calculation
SensorIVSvc::read                          INFO                Using measurement with voltage 700.120000 and current 0.080350 for current stability calculation
SensorIVSvc::read                          INFO                Using measurement with voltage 700.120000 and current 0.080350 for current stability calculation
SensorIVSvc::CalculateVBD                  INFO                Average voltage 5.000000, current 0.016475, dIdV 0.000384
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.116562 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 14.970000, current 0.023450, dIdV 0.001012
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.646025 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 24.990000, current 0.032045, dIdV 0.000704
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.548690 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 35.009998, current 0.036690, dIdV 0.000224
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.213317 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 45.029999, current 0.039035, dIdV 0.000245
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.282063 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 55.040001, current 0.041580, dIdV 0.000264
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.349460 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 65.050003, current 0.042175, dIdV -0.000145
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.223199 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 75.059998, current 0.041975, dIdV 0.000105
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.187762 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 85.070000, current 0.042720, dIdV 0.000044
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.087444 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 95.070000, current 0.043715, dIdV 0.000155
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.337765 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 105.059998, current 0.044310, dIdV -0.000036
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.085357 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 115.070000, current 0.044405, dIdV 0.000055
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.142241 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 125.070000, current 0.044005, dIdV -0.000135
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.384463 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 135.069992, current 0.043750, dIdV 0.000084
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.258817 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 145.089996, current 0.044395, dIdV 0.000045
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.146774 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 155.080002, current 0.044040, dIdV -0.000116
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.410116 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 165.069992, current 0.044335, dIdV 0.000175
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.650267 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 175.080002, current 0.045685, dIdV 0.000095
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.364071 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 185.080002, current 0.046130, dIdV -0.000006
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.024073 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 195.089996, current 0.047525, dIdV 0.000284
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.167589 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 205.089996, current 0.048220, dIdV -0.000146
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.622214 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 215.089996, current 0.048715, dIdV 0.000245
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.079583 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 225.110001, current 0.050810, dIdV 0.000174
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.769356 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 235.130005, current 0.052105, dIdV 0.000085
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.382807 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 245.160004, current 0.053205, dIdV 0.000134
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.619580 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 255.160004, current 0.054650, dIdV 0.000155
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.721911 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 265.160004, current 0.055795, dIdV 0.000075
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.355010 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 275.190002, current 0.056840, dIdV 0.000134
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.647464 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 285.200012, current 0.057885, dIdV 0.000075
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.369526 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 295.059998, current 0.058035, dIdV -0.000046
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.235378 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 304.920013, current 0.058535, dIdV 0.000145
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.755333 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 314.940002, current 0.059380, dIdV 0.000024
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.126784 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 324.970001, current 0.059825, dIdV 0.000065
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.352376 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 334.959991, current 0.060425, dIdV 0.000055
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.306111 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 344.950012, current 0.060070, dIdV -0.000126
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.722107 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 354.959991, current 0.060065, dIdV 0.000125
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.738700 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 364.970001, current 0.061060, dIdV 0.000074
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.441433 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 374.959991, current 0.061755, dIdV 0.000065
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.396248 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 384.950012, current 0.061705, dIdV -0.000075
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.466958 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 394.970001, current 0.063050, dIdV 0.000343
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 2.150650 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 404.980011, current 0.064695, dIdV -0.000015
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.093898 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 414.959991, current 0.065340, dIdV 0.000145
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.918185 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 424.970001, current 0.066185, dIdV 0.000025
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.159566 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 434.980011, current 0.067135, dIdV 0.000166
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.073359 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 444.970001, current 0.067580, dIdV -0.000076
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.499411 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 454.959991, current 0.066725, dIdV -0.000095
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.650353 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 464.929993, current 0.067120, dIdV 0.000174
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.207687 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 474.940002, current 0.067615, dIdV -0.000075
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.524715 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 484.959991, current 0.067115, dIdV -0.000025
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.180645 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 494.970001, current 0.067310, dIdV 0.000064
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.469690 and VBD limit 5.500000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 504.980011, current 0.067605, dIdV -0.000005
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.037348 and VBD limit 5.549800
SensorIVSvc::CalculateVBD                  INFO                Average voltage 514.970032, current 0.069400, dIdV 0.000365
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 2.706409 and VBD limit 5.649700
SensorIVSvc::CalculateVBD                  INFO                Average voltage 524.970032, current 0.071495, dIdV 0.000055
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.403045 and VBD limit 5.749700
SensorIVSvc::CalculateVBD                  INFO                Average voltage 534.979980, current 0.071390, dIdV -0.000076
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -0.569526 and VBD limit 5.849800
SensorIVSvc::CalculateVBD                  INFO                Average voltage 544.989990, current 0.072135, dIdV 0.000225
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.696513 and VBD limit 5.949900
SensorIVSvc::CalculateVBD                  INFO                Average voltage 555.000000, current 0.072485, dIdV -0.000155
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -1.186797 and VBD limit 6.050000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 565.010010, current 0.072030, dIdV 0.000064
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.501020 and VBD limit 6.150100
SensorIVSvc::CalculateVBD                  INFO                Average voltage 575.020020, current 0.072575, dIdV 0.000045
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.356540 and VBD limit 6.250200
SensorIVSvc::CalculateVBD                  INFO                Average voltage 585.030029, current 0.073870, dIdV 0.000214
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.691438 and VBD limit 6.350300
SensorIVSvc::CalculateVBD                  INFO                Average voltage 595.039978, current 0.074065, dIdV -0.000175
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -1.405954 and VBD limit 6.450400
SensorIVSvc::CalculateVBD                  INFO                Average voltage 605.049988, current 0.074660, dIdV 0.000293
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 2.377841 and VBD limit 6.550500
SensorIVSvc::CalculateVBD                  INFO                Average voltage 615.070007, current 0.076905, dIdV 0.000155
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.237183 and VBD limit 6.650700
SensorIVSvc::CalculateVBD                  INFO                Average voltage 625.080017, current 0.076955, dIdV -0.000145
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD -1.177787 and VBD limit 6.750800
SensorIVSvc::CalculateVBD                  INFO                Average voltage 635.080017, current 0.077200, dIdV 0.000194
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.595927 and VBD limit 6.850800
SensorIVSvc::CalculateVBD                  INFO                Average voltage 645.090027, current 0.078945, dIdV 0.000155
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 1.264037 and VBD limit 6.950900
SensorIVSvc::CalculateVBD                  INFO                Average voltage 655.099976, current 0.080240, dIdV 0.000104
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.849083 and VBD limit 7.051000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 665.099976, current 0.081285, dIdV 0.000105
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.859144 and VBD limit 7.151000
SensorIVSvc::CalculateVBD                  INFO                Average voltage 675.109985, current 0.082030, dIdV 0.000044
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.361399 and VBD limit 7.251100
SensorIVSvc::CalculateVBD                  INFO                Average voltage 685.109985, current 0.082525, dIdV 0.000055
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.457517 and VBD limit 7.351100
SensorIVSvc::CalculateVBD                  INFO                Average voltage 695.099976, current 0.082875, dIdV 0.000015
SensorIVSvc::CalculateVBD                  INFO                V_avg_BD 0.125810 and VBD limit 7.451000
SensorIVSvc::CalculateVBD                  INFO                Calculated VBD to be 690.099976
SensorIVSvc::read                          INFO                Mean voltage for current stability 0.081275 calculated with 4 points
SensorIVSvc::read                          INFO                RMS current stability 0.013119
IService::UPLOADDB                         INFO                Working on test code MODULE_IV_PS_V1
SensorIVSvc::uploadDB                      INFO                Formatting database upload for MODULE_IV_PS_V1 and run number 1
SensorIVSvc::writeOutput                   INFO                Writing results output file /home/data/output//SensorIV_PostTabIV_PPB_v1//SCIPP-PRO_LS-0298/1//20USBML1236957-HV_TAB_ATTACHED_IV_001.dat
SensorIVSvc::getDate                       INFO                Parsing date 2025-11-14T18:10:44.001+00:00...
SensorIVSvc::getDate                       INFO                Year 2025 
SensorIVSvc::getDate                       INFO                Month 11  
SensorIVSvc::getDate                       INFO                Day 14    
SensorIVSvc::getDate                       INFO                Hour 18   
SensorIVSvc::getDate                       INFO                Minute 10 
SensorIVSvc::getDate                       INFO                Second 44 
SensorIVSvc::getDate                       INFO                Offset 0:0
SensorIVSvc::getDate                       INFO                Converted month into string Nov
SensorIVSvc::getDate                       INFO                FINALLY obtained 14 Nov 2025
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
SensorIVSvc::uploadDB                      INFO                Formatting database upload for VISUAL_INSPECTION and run number 1
IService::log                              INFO                Logging status 0111 for component SCIPP-PRO_LS-0298
IService::install                          INFO                Installing output directory to /home/data/modules/SCIPP-PRO_LS-0298/SensorIV_PostTabIV_PPB_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed IVCurve.pdf...
IService::install                          INFO                Installed 20USBML1236957-HV_TAB_ATTACHED_IV_001.dat...
IService::install                          INFO                Installed DB_READY_MODULE_MODULE_IV_PS_V1.json...
IService::install                          INFO                Installed DB_READY_MODULE_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed SCIPP-PRO_LS-0298.root...
IService::install                          INFO                Installed status.json...
TopControl::monitor                        INFO                Service with algoirthm name ElectricalIV ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_Module_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_AMACIV_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SixWayJig_Y_slots ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPanelMetrology ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPanelMetrology_v7 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology_Y_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_v10 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_v10-beta ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridToSensorMetrology_v2a ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_v8a ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPA_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_SS_wuerth_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleLS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleSS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PROD_v2 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalModules_PPA_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name TCModules_PROD_v0 ignored through users request!
MetadataValue
Component type ATLAS18LS
Operator FMM
Date 2025-11-14T18:10:44.001+00:00
Institute UCSC
Power supply CAEN_DT1415ET
Temperarture [C] 19.5
Humidity 9.5
Current per area @500V 0.739
Breakdown voltage 690.100
Current stability (RMS@700V) 0.013

The current measurements are calibrated to a common temperature using the following relationship: $$I_{temp}^{SF} = \left ( \frac{T_{ref}}{T_{meas}} \right )^{2} \times e^{eE_{g}^{Si} \frac{T_{ref}-T_{meas}}{2 k_{B} T_{ref} T_{meas} }} $$ where \( T_{ref} \) is the reference temperature, \( T_{meas} \) is the temperatured measured during the test, \( e \) is the electric charge, \( E_{g}^{Si} \) is the band gap energy of silicon, and \( k_{B} \) is the Boltzman constant

Glue weight for SCIPP-PRO_LS-0298


Shown below are results for the glue weight measurements

IService::UPLOADDB                         INFO                Working on test code GLUE_WEIGHT
GlueWeightSvc::uploadDB                    INFO                Formatting database test for GLUE_WEIGHT
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
IService::log                              INFO                Logging status 0111 for component SCIPP-PRO_LS-0298
IService::install                          INFO                Installing output directory to /home/data/modules/SCIPP-PRO_LS-0298/GlueWeightModule_LS_PPA-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed DB_COMMENTS.json...
IService::install                          INFO                Installed DB_READY_MODULE_GLUE_WEIGHT.json...
IService::install                          INFO                Installed DB_READY_MODULE_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed SCIPP-PRO_LS-0298.root...
IService::install                          INFO                Installed status.json...
TopControl::monitor                        INFO                Service with algoirthm name GlueWeightModule_SS_PPA ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridBonding_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleBonding_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPullTest ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PullTestCicorelCard ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridFlexArray_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name Powerboards_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ABC_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ABC_Wafer_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HCC_PullTest_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorBondingQC_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HVTabSheet_SheerTest_Before_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HVTabSheet_SheerTest_After_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PhotoUpload ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PictureModule ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name GeneralPictures ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name PreHVTabIV ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_PreTabIV_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name Module_postHVTabIV_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_PostTabIV_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalIV ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_Module_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SensorIV_AMACIV_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name SixWayJig_Y_slots ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPanelMetrology ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridPanelMetrology_v7 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridMetrology_Y_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_v10 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_v10-beta ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_Y_v0_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyHybrid_PROD_X_v2_thick_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name HybridToSensorMetrology_v2a ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_v8 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_v8a ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPA_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_SS_wuerth_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PPB_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleLS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleSS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PROD_v2 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalModules_PPA_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name TCModules_PROD_v0 ignored through users request!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!

MetadataValue
Operator FMM, KA
Date 2025-11-18T19:19:30.671+00:00
X-hybrid program LS_Hybrid.src
Powerboard program LS_PB.src
Gluing method Dispenser
FieldWeight/Volume
Weight bare sensor [g] 7.163
Weight of x-hybrid [g] 3.325
Weight of x-tabs [g] 0.209
Weight of back-end wirebonds 0.000
Weights of powerboard [g] 2.306
Weight of stuffed module [g] 12.836
Weight of glue [mg] 250.900

Metrology results


Shown below are the results of the module metrology scans

MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_0] Height in z :                               679.696167 +/- 21.919463 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_0] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             679.873629
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 69.873629 +/- 24.125272 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 680.873629 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_1] Height in z :                               710.945667 +/- 8.776114 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_1] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             711.123129
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 101.123129 +/- 13.363610 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 712.123129 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_2] Height in z :                               711.433500 +/- 8.134347 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_2] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             711.610963
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 101.610963 +/- 12.951197 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 712.610963 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_3] Height in z :                               721.399000 +/- 5.058196 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_3] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             721.576463
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 111.576463 +/- 11.276136 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 722.576463 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_4] Height in z :                               725.243167 +/- 3.930615 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_4] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             725.420629
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 115.420629 +/- 10.817376 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 726.420629 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_5] Height in z :                               726.377833 +/- 2.253939 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_5] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             726.555296
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 116.555296 +/- 10.326962 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 727.555296 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_6] Height in z :                               724.466000 +/- 6.039217 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_6] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             724.643463
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 114.643463 +/- 11.748959 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 725.643463 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_7] Height in z :                               720.854167 +/- 6.485228 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_7] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             721.031629
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 111.031629 +/- 11.984327 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 722.031629 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_8] Height in z :                               723.963333 +/- 7.560146 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_8] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             724.140796
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 114.140796 +/- 12.598480 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 725.140796 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_9] Height in z :                               707.636000 +/- 14.223779 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [ABC_X_9] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             707.813463
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 97.813463 +/- 17.432205 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 708.813463 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [HCC_X_0] Height in z :                               725.954333 +/- 9.952653 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [HCC_X_0] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             726.131796
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 116.131796 +/- 14.164081 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 727.131796 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [H_X_0] Height in z :                               660.480000 +/- 13.864201 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [H_X_0] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             660.657463
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 50.657463 +/- 17.140069 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 661.657463 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [H_X_1] Height in z :                               637.439750 +/- 6.531000 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [H_X_1] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             637.617213
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 27.617213 +/- 12.009157 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 638.617213 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [H_X_2] Height in z :                               657.211500 +/- 34.458374 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [H_X_2] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             657.388963
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 610.000000 +/- 10.000000 extra material : 47.388963 +/- 35.901886 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 658.388963 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_1] Height in z :                               675.486600 +/- 14.232125 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_1] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             675.664062
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 545.000000 +/- 15.000000 extra material : 130.664062 +/- 20.715195 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 676.664062 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_2] Height in z :                               605.824000 +/- 0.000000 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_2] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             606.001463
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 545.000000 +/- 15.000000 extra material : 61.001463 +/- 15.052106 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 607.001463 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_3] Height in z :                               705.837000 +/- 5.991500 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_3] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             706.014463
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 545.000000 +/- 15.000000 extra material : 161.014463 +/- 16.200740 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 707.014463 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_4] Height in z :                               670.358750 +/- 7.563250 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_4] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             670.536213
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 545.000000 +/- 15.000000 extra material : 125.536213 +/- 16.845434 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 671.536213 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeGlueHeights           INFO                Printing associated surfaces...
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_5] Height in z :                               525.662000 +/- 9.136527 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> [PB_5] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Difference in heights :                             525.839463
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue assuming 545.000000 +/- 15.000000 extra material : -19.160537 +/- 17.608010 um 
MetrologySvc::analyzeGlueHeights           INFO                 ----> Thickeness of glue using derived value -1.000000 from hybrid measurements : 526.839463 um 
MetrologySvc::analyzeGlueHeights           INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                Printing associated surfaces for total height
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C1] Height in z :                               3116.235000 +/- 0.000000 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C1] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> Difference in heights :                             3116.412462
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                Printing associated surfaces for total height
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C2] Height in z :                               3055.263000 +/- 0.000000 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C2] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> Difference in heights :                             3055.440462
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                Printing associated surfaces for total height
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C3] Height in z :                               3130.596000 +/- 0.000000 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C3] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> Difference in heights :                             3130.773462
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                Printing associated surfaces for total height
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C4] Height in z :                               3133.850000 +/- 0.000000 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> [C4] Height in z :                               -0.177463 +/- 1.251360 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> Difference in heights :                             3134.027462
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::analyzeTotalHeights          INFO                Printing associated surfaces for total height
MetrologySvc::analyzeTotalHeights          INFO                 ----> [Shield] Height in z :                               5754.675200 +/- 41.503434 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> [Shield] Height in z :                               15.652837 +/- 37.611278 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> Difference in heights :                             5739.022362
MetrologySvc::analyzeTotalHeights          INFO                 ----> [Shield] Max height in z (shieldbox) :                               5852.576000 +/- 41.503434 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> [Shield] Height in z :                               15.652837 +/- 37.611278 um 
MetrologySvc::analyzeTotalHeights          INFO                 ----> Max difference in heights (shieldbox) :                             5836.923163
MetrologySvc::analyzeTotalHeights          INFO                --------------------------------------------------------------------------------- 
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
IService::UPLOADDB                         INFO                Working on test code MODULE_METROLOGY
MetrologySvc::uploadDB                     INFO                Doing DB upload...
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyModules_PROD_LS_v1_interposer//SCIPP-PRO_LS-0298/1//Data__SCIPP-PRO_LS-0298.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyModules_PROD_LS_v1_interposer//SCIPP-PRO_LS-0298/1//Results__SCIPP-PRO_LS-0298.txt
MetrologySvc::writeOutput                  INFO                Writing common output file /home/data/output//MetrologyModules_PROD_LS_v1_interposer//SCIPP-PRO_LS-0298/1//Data__SCIPP-PRO_LS-0298.txt
MetrologySvc::writeOutput                  INFO                Writing results output file /home/data/output//MetrologyModules_PROD_LS_v1_interposer//SCIPP-PRO_LS-0298/1//Results__SCIPP-PRO_LS-0298.txt
IService::log                              INFO                Logging status 0111 for component SCIPP-PRO_LS-0298
IService::install                          INFO                Installing output directory to /home/data/modules/SCIPP-PRO_LS-0298/MetrologyModules_PROD_LS_v1_interposer-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed Summary.pdf...
IService::install                          INFO                Installed TotalStackUp.pdf...
IService::install                          INFO                Installed Positions.pdf...
IService::install                          INFO                Installed Data__SCIPP-PRO_LS-0298.txt...
IService::install                          INFO                Installed Results__SCIPP-PRO_LS-0298.txt...
IService::install                          INFO                Installed Results.json...
IService::install                          INFO                Installed DB_READY_MODULE_MODULE_METROLOGY.json...
IService::install                          INFO                Installed SCIPP-PRO_LS-0298.root...
IService::install                          INFO                Installed status.json...
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_LS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name MetrologyModules_PROD_SS_v1_thin_interposer ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleLS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ModuleSS_Bow_PPA_v0 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PPB_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name BurnInHybrids_PROD_v2 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name ElectricalModules_PPA_v1 ignored through users request!
TopControl::monitor                        INFO                Service with algoirthm name TCModules_PROD_v0 ignored through users request!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
MetadataValue
Component type
Operator AS
Date 2025-11-18T18:17:48.773+00:00
Program version PROD_v1
Instrument SmartScope Flash 250

The glue thicknesses are shown below.


The positions are defined as the hybrid cross fiducials and the circles on the powerboard

.

Height of components on the powerboard

.

Wirebonding for SCIPP-PRO_LS-0298


Wirebonding summaries will be given below

WireBondSvc::run                           INFO                Loading group FAILED_HYBRID_TO_PB
WireBondSvc::run                           INFO                Loading group FAILED_MODULE_TO_FRAME
WireBondSvc::run                           INFO                Loading group FAILED_ROW_1
WireBondSvc::run                           INFO                Loading group FAILED_ROW_2
WireBondSvc::run                           INFO                Loading group FAILED_ROW_3
WireBondSvc::run                           INFO                Loading group FAILED_ROW_4
WireBondSvc::run                           INFO                Loading group REPAIRED_HYBRID_TO_PB
WireBondSvc::run                           INFO                Loading group REPAIRED_MODULE_TO_FRAME
WireBondSvc::run                           INFO                Loading group REPAIRED_ROW_1
WireBondSvc::run                           INFO                Loading group REPAIRED_ROW_2
WireBondSvc::run                           INFO                Loading group REPAIRED_ROW_3
WireBondSvc::run                           INFO                Loading group REPAIRED_ROW_4
IService::UPLOADDB                         INFO                Working on test code MODULE_WIRE_BONDING
IService::UPLOADDB                         INFO                Working on test code VISUAL_INSPECTION
IService::log                              INFO                Logging status 0111 for component SCIPP-PRO_LS-0298
IService::install                          INFO                Installing output directory to /home/data/modules/SCIPP-PRO_LS-0298/ModuleBonding_v1-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed DB_READY_MODULE_MODULE_WIRE_BONDING.json...
IService::install                          INFO                Installed DB_READY_MODULE_VISUAL_INSPECTION.json...
IService::install                          INFO                Installed SCIPP-PRO_LS-0298.root...
IService::install                          INFO                Installed status.json...
MetaData::readField                        INFO                Reading file /home/data/upload//ModuleBonding_v1//SCIPP-PRO_LS-0299-1.config
IService::next                             INFO                Scheduling ModuleBonding_SCIPP-PRO_LS-0299-1.json file to run...
IService::next                             INFO                Writing results to /home/data/output//ModuleBonding_v1//SCIPP-PRO_LS-0299/1/
MetadataValue
Date 2025-11-21T00:26:39.134+00:00
Operator LM
Bonder hk2
Version v1
Bond numberTypeComment

Thermal cycling results


Thermal cycling summaries will be given below

TCSvc::getListOfFilesFromDir               INFO                Scanning for test files 501-0661/20USBP05010661
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files 501-0661/20USBP05010661
TCSvc::read                                INFO                Found 0 results files...
TCSvc::getListOfFilesFromDir               INFO                Scanning for test files GPC2328_X_019_B_H1/20USBHX2003449
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files GPC2328_X_019_B_H1/20USBHX2003449
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251203_11_3PG_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251203_11_NO_TC.json
TCSvc::read                                INFO                Found 0 results files...
TCSvc::getListOfFilesFromDir               INFO                Scanning for test files VPA38892-W02331/20USBSL0002331
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files VPA38892-W02331/20USBSL0002331
TCSvc::read                                INFO                Found 0 results files...
TCSvc::getListOfFilesFromDir               INFO                Scanning for test files SCIPP-PRO_LS-0298/20USBML1236957
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files SCIPP-PRO_LS-0298/20USBML1236957
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBML1236957_20251203_11_MODULE_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::read                                INFO                Found 0 results files...
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
IService::UPLOADDB                         INFO                Working on test code PEDESTAL_TRIM_TC
IService::UPLOADDB                         INFO                Working on test code STROBE_DELAY_TC
IService::UPLOADDB                         INFO                Working on test code RESPONSE_CURVE_TC
IService::UPLOADDB                         INFO                Working on test code NO_TC
IService::UPLOADDB                         INFO                Working on test code MODULE_IV_AMAC_TC
IService::UPLOADDB                         INFO                Working on test code HVSTABILITY
IService::UPLOADDB                         INFO                Working on test code OPEN_CHANNEL_SEARCH_TC
IService::UPLOADDB                         INFO                Working on test code MODULE_TC
IService::log                              INFO                Logging status 0111 for component SCIPP-PRO_LS-0298
IService::install                          INFO                Installing output directory to /home/data/modules/SCIPP-PRO_LS-0298/TCModules_PROD_v0-1/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed 20USBHX2003449_TC_Noise.pdf...
IService::install                          INFO                Installed 20USBHX2003449_TC_Averages_RC_innse.pdf...
IService::install                          INFO                Installed 20USBHX2003449_TC_Averages_RC_gain.pdf...
IService::install                          INFO                Installed 20USBML1236957_TC_ENV_DATA.pdf...
IService::install                          INFO                Installed 20USBML1236957_TC_AMAC_IV_fast.pdf...
IService::install                          INFO                Installed 20USBML1236957_TC_AMAC_IV.pdf...
IService::install                          INFO                Installed DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json...
IService::install                          INFO                Installed DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json...
IService::install                          INFO                Installed DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_3PG_TC.json...
IService::install                          INFO                Installed DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json...
IService::install                          INFO                Installed DB_READY_NO_TC_SN20USBHX2003449_20251203_11_NO_TC.json...
IService::install                          INFO                Installed DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json...
IService::install                          INFO                Installed DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003449_20251203_11_OPEN_CHANNEL_SEARCH_TC.json...
IService::install                          INFO                Installed DB_READY_MODULE_TC_SN20USBML1236957_20251203_11_MODULE_TC.json...
IService::install                          INFO                Installed SCIPP-PRO_LS-0298.root...
IService::install                          INFO                Installed status.json...
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=70.641529 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=96.183363 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=90.099696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=95.068029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=109.915196 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=108.321029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=107.921863 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=105.885029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=100.054696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=83.079696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 96.717012 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=160.795862 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=33.994863 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=173.612362 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=157.778613 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 131.545425 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5758.805562 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5758.805562 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=70.641529 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=96.183363 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=90.099696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=95.068029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=109.915196 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=108.321029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=107.921863 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=105.885029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=100.054696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=83.079696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 96.717012 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=160.795862 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=33.994863 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=173.612362 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=157.778613 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 131.545425 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5758.805562 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5758.805562 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=70.641529 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=96.183363 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=90.099696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=95.068029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=109.915196 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=108.321029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=107.921863 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=105.885029 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=100.054696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=83.079696 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 96.717012 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=160.795862 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=33.994863 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=173.612362 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=157.778613 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 131.545425 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5758.805562 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5758.805562 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003452_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003452_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003452_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003452_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003452_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236960_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003452_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236960_20251203_11_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003452_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003452_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003452_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003452_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003452_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236960_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003452_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236960_20251203_11_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003452_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003452_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003452_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003452_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003452_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236960_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003452_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236960_20251203_11_MODULE_TC.json
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=80.624250 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=97.023917 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.685583 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=114.780750 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=116.251917 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=113.587250 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=110.546250 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=113.994417 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.443083 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=95.144583 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.808200 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=123.511850 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=84.700250 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=129.436750 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=107.120500 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 111.192338 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5819.383250 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5819.383250 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003461_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003461_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003461_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003461_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003461_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236961_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003461_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236961_20251203_11_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003461_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003461_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003461_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003461_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003461_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236961_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003461_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236961_20251203_11_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003461_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003461_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003461_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003461_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003461_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236961_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003461_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236961_20251203_11_MODULE_TC.json
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=77.920588 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=107.199587 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=111.068754 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=109.874421 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.166421 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=117.115754 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=120.511087 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=121.673087 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=123.146754 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=78.536754 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 108.221321 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=128.035488 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=131.416088 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=168.833087 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=130.350838 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 139.658875 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5777.095525 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5777.095525 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003745_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003745_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003745_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003745_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003745_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236963_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003745_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236963_20251203_11_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003745_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003745_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003745_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003745_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003745_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236963_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003745_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236963_20251203_11_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003745_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003745_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003745_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003745_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003745_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236963_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003745_20251203_11_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236963_20251203_11_MODULE_TC.json
TCSvc::PassFail                            WARNING             Failing test detected in DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            WARNING             Failing test detected in DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            WARNING             Failing test detected in DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
  • ✔ Component is at UCSC. OK to upload
  • ✘ Component fails stage checks. Current stage is FINISHED but requires one of { "BONDED", "TESTED" }. Prohibiting upload!
  • ✘ Component has failed criteria!

Test Summary

TestPassedProblems
Open Channel Search True False
Pedestal Trim True False
Strobe Delay True False
3PG True False
Response Curve True False
Noise Occupancy True False
AMAC IV False False
HV Stability MISSING! MISSING!
Module TC True False

Response Curve Defect Summary

Defect #DescriptionChip BankRun NumberChip #Channel(s)
1 11-10 Strip with high noise away 3 456
2 11-10 Strip with high gain away 9 1279
3 11-10 Strip with high gain under 1 128
4 11-115 Strip with high noise away 0 0
5 11-115 Strip with high gain away 9 1279
6 11-115 Strip with low gain under 7 910
7 11-132 Strip with high gain away 9 1279
8 11-132 Strip with high gain under 1 128
9 11-149 Strip with high gain away 9 1279
10 11-149 Strip with high noise under 5 684
11 11-149 Strip with low gain under 7 910
12 11-166 Strip with high gain away 9 1279
13 11-166 Strip with high gain under 1 128
14 11-183 Strip with high gain away 9 1279
15 11-183 Strip with low gain under 7 910
16 11-200 Strip with high noise away 0 68
17 11-200 Strip with high gain away 9 1279
18 11-200 Strip with high gain under 1 128
19 11-200 Strip with high noise under 1 196
20 11-217 Strip with high noise away 0 80
21 11-217 Strip with high gain away 9 1279
22 11-217 Strip with low gain under 0 20
23 11-217 Strip with low gain under 7 910
24 11-234 Strip with high gain away 9 1279
25 11-234 Strip with high gain under 1 128
26 11-234 Strip with low gain under 7 910
27 11-251 Strip with high noise away 0 0
28 11-251 Strip with high gain away 9 1279
29 11-251 Strip with low gain under 7 910
30 11-268 Strip with high gain away 9 1279
31 11-268 Strip with high gain under 1 128
32 11-285 Strip with high noise away 1 221
33 11-285 Strip with high gain away 9 1279
34 11-285 Strip with high noise under 1 224
35 11-285 Strip with low gain under 7 910
36 11-302 Strip with high gain away 9 1279
37 11-302 Strip with high gain under 1 128
38 11-31 Strip with high noise away 5 704
39 11-31 Strip with high gain away 9 1279
40 11-31 Strip with low gain under 7 910
41 11-319 Strip with high noise away 0 0
42 11-319 Strip with high gain away 9 1279
43 11-319 Strip with high noise under 5 726
44 11-319 Strip with low gain under 7 910
45 11-336 Strip with high gain away 9 1279
46 11-336 Strip with high gain under 1 128
47 11-336 Strip with high noise under 1 196
48 11-353 Strip with high noise away 8 1030
49 11-353 Strip with high gain away 9 1279
50 11-353 Strip with high noise under 5 684
51 11-353 Strip with low gain under 7 910
52 11-370 Strip with high gain away 9 1279
53 11-370 Strip with high gain under 1 128
54 11-370 Strip with low gain under 7 910
55 11-388 Strip with high noise away 3 443
56 11-388 Strip with high gain away 9 1279
57 11-388 Strip with high gain under 1 128
58 11-388 Strip with low gain under 7 910
59 11-404 Strip with high gain away 9 1279
60 11-404 Strip with high noise under 5 726
61 11-404 Strip with high noise under 8 1099
62 11-421 Strip with high gain away 9 1279
63 11-421 Strip with high gain under 1 128
64 11-421 Strip with high noise under 6 780
65 11-47 Strip with high gain away 9 1279
66 11-47 Strip with low gain under 7 910
67 11-64 Strip with high gain away 9 1279
68 11-64 Strip with high gain under 1 128
69 11-64 Strip with high noise under 5 671
70 11-81 Strip with high noise away 5 704
71 11-81 Strip with high gain away 9 1279
72 11-81 Strip with low gain under 7 910
73 11-98 Strip with high gain away 9 1279
74 11-98 Strip with high gain under 1 128

Comments

Comments
11-10 - 3 defects found affecting 3 strips 11-10 - 1 maximum consecutive defects 11-10 - Cuts from: LS link 0 11-10 - Cuts from: LS link 1 11-31 - 3 defects found affecting 3 strips 11-31 - 1 maximum consecutive defects 11-31 - Cuts from: LS link 0 11-31 - Cuts from: LS link 1 11-47 - 2 defects found affecting 2 strips 11-47 - 1 maximum consecutive defects 11-47 - Cuts from: LS link 0 11-47 - Cuts from: LS link 1 11-64 - 3 defects found affecting 3 strips 11-64 - 1 maximum consecutive defects 11-64 - Cuts from: LS link 0 11-64 - Cuts from: LS link 1 11-81 - 3 defects found affecting 3 strips 11-81 - 1 maximum consecutive defects 11-81 - Cuts from: LS link 0 11-81 - Cuts from: LS link 1 11-98 - 2 defects found affecting 2 strips 11-98 - 1 maximum consecutive defects 11-98 - Cuts from: LS link 0 11-98 - Cuts from: LS link 1 11-115 - 3 defects found affecting 3 strips 11-115 - 1 maximum consecutive defects 11-115 - Cuts from: LS link 0 11-115 - Cuts from: LS link 1 11-132 - 2 defects found affecting 2 strips 11-132 - 1 maximum consecutive defects 11-132 - Cuts from: LS link 0 11-132 - Cuts from: LS link 1 11-149 - 3 defects found affecting 3 strips 11-149 - 1 maximum consecutive defects 11-149 - Cuts from: LS link 0 11-149 - Cuts from: LS link 1 11-166 - 2 defects found affecting 2 strips 11-166 - 1 maximum consecutive defects 11-166 - Cuts from: LS link 0 11-166 - Cuts from: LS link 1 11-183 - 2 defects found affecting 2 strips 11-183 - 1 maximum consecutive defects 11-183 - Cuts from: LS link 0 11-183 - Cuts from: LS link 1 11-200 - 4 defects found affecting 4 strips 11-200 - 1 maximum consecutive defects 11-200 - Cuts from: LS link 0 11-200 - Cuts from: LS link 1 11-217 - 4 defects found affecting 4 strips 11-217 - 1 maximum consecutive defects 11-217 - Cuts from: LS link 0 11-217 - Cuts from: LS link 1 11-234 - 3 defects found affecting 3 strips 11-234 - 1 maximum consecutive defects 11-234 - Cuts from: LS link 0 11-234 - Cuts from: LS link 1 11-251 - 3 defects found affecting 3 strips 11-251 - 1 maximum consecutive defects 11-251 - Cuts from: LS link 0 11-251 - Cuts from: LS link 1 11-268 - 2 defects found affecting 2 strips 11-268 - 1 maximum consecutive defects 11-268 - Cuts from: LS link 0 11-268 - Cuts from: LS link 1 11-285 - 4 defects found affecting 4 strips 11-285 - 1 maximum consecutive defects 11-285 - Cuts from: LS link 0 11-285 - Cuts from: LS link 1 11-302 - 2 defects found affecting 2 strips 11-302 - 1 maximum consecutive defects 11-302 - Cuts from: LS link 0 11-302 - Cuts from: LS link 1 11-319 - 4 defects found affecting 4 strips 11-319 - 1 maximum consecutive defects 11-319 - Cuts from: LS link 0 11-319 - Cuts from: LS link 1 11-336 - 3 defects found affecting 3 strips 11-336 - 1 maximum consecutive defects 11-336 - Cuts from: LS link 0 11-336 - Cuts from: LS link 1 11-353 - 4 defects found affecting 4 strips 11-353 - 1 maximum consecutive defects 11-353 - Cuts from: LS link 0 11-353 - Cuts from: LS link 1 11-370 - 3 defects found affecting 3 strips 11-370 - 1 maximum consecutive defects 11-370 - Cuts from: LS link 0 11-370 - Cuts from: LS link 1 11-388 - 4 defects found affecting 4 strips 11-388 - 1 maximum consecutive defects 11-388 - Cuts from: LS link 0 11-388 - Cuts from: LS link 1 11-404 - 3 defects found affecting 3 strips 11-404 - 1 maximum consecutive defects 11-404 - Cuts from: LS link 0 11-404 - Cuts from: LS link 1 11-421 - 3 defects found affecting 3 strips 11-421 - 1 maximum consecutive defects 11-421 - Cuts from: LS link 0 11-421 - Cuts from: LS link 1

Figures of gain and noise, and correlations with wafer/hybrid level test results, if they exist.

Thermal cycling results


Thermal cycling summaries will be given below

TCSvc::getListOfFilesFromDir               INFO                Scanning for test files 501-0661/20USBP05010661
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files 501-0661/20USBP05010661
TCSvc::read                                INFO                Found 0 results files...
TCSvc::getListOfFilesFromDir               INFO                Scanning for test files GPC2328_X_019_B_H1/20USBHX2003449
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files GPC2328_X_019_B_H1/20USBHX2003449
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251205_2_STROBE_DELAY_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251205_2_NO_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBHX2003449_20251205_2_3PG_TC.json
TCSvc::read                                INFO                Found 0 results files...
TCSvc::getListOfFilesFromDir               INFO                Scanning for test files VPA38892-W02331/20USBSL0002331
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files VPA38892-W02331/20USBSL0002331
TCSvc::read                                INFO                Found 0 results files...
TCSvc::getListOfFilesFromDir               INFO                Scanning for test files SCIPP-PRO_LS-0298/20USBML1236957
TCSvc::getListOfJsonFilesFromDir           INFO                Scanning for test files SCIPP-PRO_LS-0298/20USBML1236957
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBML1236957_20251205_2_MODULE_TC.json
TCSvc::getListOfJsonFilesFromDir           INFO                Loading json fle SN20USBML1236957_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::read                                INFO                Found 0 results files...
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
IService::UPLOADDB                         INFO                Working on test code PEDESTAL_TRIM_TC
IService::UPLOADDB                         INFO                Working on test code STROBE_DELAY_TC
IService::UPLOADDB                         INFO                Working on test code RESPONSE_CURVE_TC
IService::UPLOADDB                         INFO                Working on test code NO_TC
IService::UPLOADDB                         INFO                Working on test code MODULE_IV_AMAC_TC
IService::UPLOADDB                         INFO                Working on test code HVSTABILITY
IService::UPLOADDB                         INFO                Working on test code OPEN_CHANNEL_SEARCH_TC
IService::UPLOADDB                         INFO                Working on test code MODULE_TC
IService::log                              INFO                Logging status 0111 for component SCIPP-PRO_LS-0298
IService::install                          INFO                Installing output directory to /home/data/modules/SCIPP-PRO_LS-0298/TCModules_PROD_v0-2/
IService::install                          INFO                Installed warningLog.txt...
IService::install                          INFO                Installed errorLog.txt...
IService::install                          INFO                Installed infoLog.txt...
IService::install                          INFO                Installed 20USBHX2003449_TC_Noise.pdf...
IService::install                          INFO                Installed 20USBHX2003449_TC_Averages_RC_innse.pdf...
IService::install                          INFO                Installed 20USBHX2003449_TC_Averages_RC_gain.pdf...
IService::install                          INFO                Installed 20USBML1236957_TC_ENV_DATA.pdf...
IService::install                          INFO                Installed 20USBML1236957_TC_AMAC_IV_fast.pdf...
IService::install                          INFO                Installed 20USBML1236957_TC_AMAC_IV.pdf...
IService::install                          INFO                Installed DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251205_2_PEDESTAL_TRIM_TC.json...
IService::install                          INFO                Installed DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251205_2_STROBE_DELAY_TC.json...
IService::install                          INFO                Installed DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_RESPONSE_CURVE_TC.json...
IService::install                          INFO                Installed DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_3PG_TC.json...
IService::install                          INFO                Installed DB_READY_NO_TC_SN20USBHX2003449_20251205_2_NO_TC.json...
IService::install                          INFO                Installed DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251205_2_MODULE_IV_AMAC_TC.json...
IService::install                          INFO                Installed DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003449_20251205_2_OPEN_CHANNEL_SEARCH_TC.json...
IService::install                          INFO                Installed DB_READY_MODULE_TC_SN20USBML1236957_20251205_2_MODULE_TC.json...
IService::install                          INFO                Installed SCIPP-PRO_LS-0298.root...
IService::install                          INFO                Installed status.json...
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=69.873629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=101.123129 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=101.610963 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=111.576463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=115.420629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=116.555296 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=114.643463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=111.031629 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=114.140796 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=97.813463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 105.378946 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=130.664062 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=61.001463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=161.014463 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=125.536213 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 119.554050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5836.923163 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5836.923163 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251203_11_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251203_11_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251203_11_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251203_11_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003449_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236957_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003449_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236957_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003449_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003449_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003449_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003449_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003449_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236957_20251205_2_MODULE_TC.json
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=116.807967 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=142.444467 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=138.025800 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=142.605300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=153.544300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=152.747967 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=156.961300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=163.200800 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=164.518467 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=141.591633 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 147.244800 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=96.989600 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=118.933800 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=132.250550 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=86.642300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 108.704063 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5795.419638 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5795.419638 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=116.807967 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=142.444467 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=138.025800 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=142.605300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=153.544300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=152.747967 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=156.961300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=163.200800 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=164.518467 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=141.591633 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 147.244800 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=96.989600 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=118.933800 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=132.250550 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=86.642300 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 108.704063 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5795.419638 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5795.419638 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003764_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003764_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003764_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003764_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003764_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236962_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003764_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236962_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003764_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003764_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003764_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003764_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003764_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236962_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003764_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236962_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003764_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003764_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003764_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003764_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003764_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236962_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003764_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236962_20251205_2_MODULE_TC.json
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=107.198033 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=135.719367 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=132.798367 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=138.743367 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=144.864867 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=148.142367 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=144.635533 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=132.154367 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=132.827700 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=103.880033 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 132.096400 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=89.225200 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=123.779200 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=112.423450 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=110.422950 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 108.962700 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5821.815925 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5821.815925 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003766_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003766_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003766_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003766_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003766_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236964_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003766_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236964_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003766_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003766_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003766_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003766_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003766_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236964_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003766_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236964_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003766_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003766_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003766_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003766_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003766_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236964_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003766_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236964_20251205_2_MODULE_TC.json
MetrologySvc::PassFail_Module              INFO                Thickness group ABC_X
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_0 point with z=127.019533 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_1 point with z=140.619700 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_2 point with z=139.226200 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_3 point with z=150.699033 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_4 point with z=150.301867 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_5 point with z=145.945367 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_6 point with z=144.422533 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_7 point with z=146.784700 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_8 point with z=151.560033 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement ABC_X_9 point with z=113.871533 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 141.045050 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group PB
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 170.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 40.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 70.000000
MetrologySvc::PassFail_Module              INFO                Including measurement PB_1 point with z=110.534800 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_2 point with z=150.623200 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_3 point with z=127.517700 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Including measurement PB_4 point with z=101.641450 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 122.579287 is within allowed range!
MetrologySvc::PassFail_Module              INFO                Thickness group ShieldBox
MetrologySvc::PassFail_Module              INFO                 >>> Minimum allowed 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum allowed 6110.000000
MetrologySvc::PassFail_Module              INFO                 >>> Minimum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                 >>> Maximum with problems 0.000000
MetrologySvc::PassFail_Module              INFO                Including measurement Shield point with z=5799.145913 for module glue height calculation
MetrologySvc::PassFail_Module              INFO                Average value 5799.145913 is within allowed range!
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003776_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003776_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003776_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003776_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003776_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236966_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003776_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236966_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003776_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003776_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003776_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003776_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003776_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236966_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003776_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236966_20251205_2_MODULE_TC.json
TCSvc::PassFail                            INFO                Reading PassFail status directly from uploaded files
TCSvc::PassFail                            INFO                Loading json fle DB_READY_PEDESTAL_TRIM_TC_SN20USBHX2003776_20251205_2_PEDESTAL_TRIM_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_STROBE_DELAY_TC_SN20USBHX2003776_20251205_2_STROBE_DELAY_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003776_20251205_2_RESPONSE_CURVE_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_RESPONSE_CURVE_TC_SN20USBHX2003776_20251205_2_3PG_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_NO_TC_SN20USBHX2003776_20251205_2_NO_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236966_20251205_2_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_OPEN_CHANNEL_SEARCH_TC_SN20USBHX2003776_20251205_2_OPEN_CHANNEL_SEARCH_TC.json
TCSvc::PassFail                            INFO                Loading json fle DB_READY_MODULE_TC_SN20USBML1236966_20251205_2_MODULE_TC.json
TCSvc::PassFail                            WARNING             Failing test detected in DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            WARNING             Failing test detected in DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json
TCSvc::PassFail                            WARNING             Failing test detected in DB_READY_MODULE_IV_AMAC_TC_SN20USBML1236957_20251203_11_MODULE_IV_AMAC_TC.json

Test Summary

TestPassedProblems
Open Channel Search True False
Pedestal Trim True False
Strobe Delay True False
3PG True False
Response Curve True False
Noise Occupancy True False
AMAC IV True False
HV Stability MISSING! MISSING!
Module TC True False

Response Curve Defect Summary

Defect #DescriptionChip BankRun NumberChip #Channel(s)
1 2-10 Strip with high noise away 1 170
2 2-10 Strip with high noise away 1 204
3 2-10 Strip with high gain away 9 1279
4 2-10 Strip with high gain under 1 128
5 2-10 Strip with high noise under 1 196
6 2-10 Strip with low gain under 7 910
7 2-115 Strip with high gain away 9 1279
8 2-115 Strip with low gain under 7 910
9 2-132 Strip with high gain away 9 1279
10 2-132 Strip with high gain under 1 128
11 2-132 Strip with high noise under 3 461
12 2-132 Strip with high noise under 5 683
13 2-149 Strip with high noise away 0 0
14 2-149 Strip with high noise away 5 704
15 2-149 Strip with high gain away 9 1279
16 2-149 Strip with high noise under 3 460
17 2-149 Strip with high noise under 5 680
18 2-149 Strip with low gain under 7 910
19 2-166 Strip with high gain away 9 1279
20 2-166 Strip with high gain under 1 128
21 2-183 Strip with high noise away 4 608
22 2-183 Strip with high gain away 9 1279
23 2-183 Strip with low gain under 7 910
24 2-200 Strip with high gain away 9 1279
25 2-200 Strip with high gain under 1 128
26 2-217 Strip with high noise away 4 608
27 2-217 Strip with high noise away 5 704
28 2-217 Strip with high noise away 6 775
29 2-217 Strip with high gain away 9 1279
30 2-217 Strip with high noise under 5 684
31 2-217 Strip with low gain under 7 910
32 2-234 Strip with high noise away 0 42
33 2-234 Strip with high gain away 9 1279
34 2-234 Strip with high gain under 1 128
35 2-234 Strip with high noise under 1 196
36 2-251 Strip with high noise away 0 0
37 2-251 Strip with high gain away 9 1279
38 2-251 Strip with low gain under 7 910
39 2-268 Strip with high gain away 9 1279
40 2-268 Strip with high gain under 1 128
41 2-285 Strip with high noise away 5 704
42 2-285 Strip with high noise away 6 820
43 2-285 Strip with high gain away 9 1279
44 2-285 Strip with high noise under 5 684
45 2-285 Strip with low gain under 7 910
46 2-302 Strip with high gain away 9 1279
47 2-302 Strip with high gain under 1 128
48 2-302 Strip with high noise under 1 226
49 2-31 Strip with high noise away 2 304
50 2-31 Strip with high noise away 4 595
51 2-31 Strip with high noise away 5 704
52 2-31 Strip with high gain away 9 1279
53 2-31 Strip with high noise under 5 684
54 2-31 Strip with low gain under 7 910
55 2-319 Strip with high noise away 0 0
56 2-319 Strip with high gain away 9 1279
57 2-319 Strip with low gain under 7 910
58 2-336 Strip with high gain away 9 1279
59 2-336 Strip with high gain under 1 128
60 2-336 Strip with low gain under 7 910
61 2-353 Strip with high noise away 5 704
62 2-353 Strip with high gain away 9 1279
63 2-353 Strip with high noise under 5 684
64 2-353 Strip with low gain under 7 910
65 2-370 Strip with high gain away 9 1279
66 2-370 Strip with high gain under 1 128
67 2-388 Strip with high noise away 1 194
68 2-388 Strip with high gain away 9 1279
69 2-388 Strip with high gain under 1 128
70 2-388 Strip with low gain under 7 910
71 2-404 Strip with high noise away 5 704
72 2-404 Strip with high gain away 9 1279
73 2-404 Strip with low gain under 7 910
74 2-421 Strip with high gain away 9 1279
75 2-421 Strip with high gain under 1 128
76 2-421 Strip with low gain under 7 910
77 2-47 Strip with high noise away 5 704
78 2-47 Strip with high gain away 9 1279
79 2-47 Strip with high gain under 1 128
80 2-47 Strip with high noise under 1 148
81 2-47 Strip with low gain under 7 910
82 2-64 Strip with high gain away 9 1279
83 2-64 Strip with high gain under 1 128
84 2-64 Strip with low gain under 7 910
85 2-81 Strip with high noise away 3 443
86 2-81 Strip with high gain away 9 1279
87 2-81 Strip with high noise under 5 684
88 2-81 Strip with low gain under 7 910
89 2-98 Strip with high gain away 9 1279
90 2-98 Strip with high gain under 1 128

Comments

Comments
2-10 - 6 defects found affecting 6 strips 2-10 - 1 maximum consecutive defects 2-10 - Cuts from: LS link 0 2-10 - Cuts from: LS link 1 2-31 - 6 defects found affecting 6 strips 2-31 - 1 maximum consecutive defects 2-31 - Cuts from: LS link 0 2-31 - Cuts from: LS link 1 2-47 - 5 defects found affecting 5 strips 2-47 - 1 maximum consecutive defects 2-47 - Cuts from: LS link 0 2-47 - Cuts from: LS link 1 2-64 - 3 defects found affecting 3 strips 2-64 - 1 maximum consecutive defects 2-64 - Cuts from: LS link 0 2-64 - Cuts from: LS link 1 2-81 - 4 defects found affecting 4 strips 2-81 - 1 maximum consecutive defects 2-81 - Cuts from: LS link 0 2-81 - Cuts from: LS link 1 2-98 - 2 defects found affecting 2 strips 2-98 - 1 maximum consecutive defects 2-98 - Cuts from: LS link 0 2-98 - Cuts from: LS link 1 2-115 - 2 defects found affecting 2 strips 2-115 - 1 maximum consecutive defects 2-115 - Cuts from: LS link 0 2-115 - Cuts from: LS link 1 2-132 - 4 defects found affecting 4 strips 2-132 - 1 maximum consecutive defects 2-132 - Cuts from: LS link 0 2-132 - Cuts from: LS link 1 2-149 - 6 defects found affecting 6 strips 2-149 - 1 maximum consecutive defects 2-149 - Cuts from: LS link 0 2-149 - Cuts from: LS link 1 2-166 - 2 defects found affecting 2 strips 2-166 - 1 maximum consecutive defects 2-166 - Cuts from: LS link 0 2-166 - Cuts from: LS link 1 2-183 - 3 defects found affecting 3 strips 2-183 - 1 maximum consecutive defects 2-183 - Cuts from: LS link 0 2-183 - Cuts from: LS link 1 2-200 - 2 defects found affecting 2 strips 2-200 - 1 maximum consecutive defects 2-200 - Cuts from: LS link 0 2-200 - Cuts from: LS link 1 2-217 - 6 defects found affecting 6 strips 2-217 - 1 maximum consecutive defects 2-217 - Cuts from: LS link 0 2-217 - Cuts from: LS link 1 2-234 - 4 defects found affecting 4 strips 2-234 - 1 maximum consecutive defects 2-234 - Cuts from: LS link 0 2-234 - Cuts from: LS link 1 2-251 - 3 defects found affecting 3 strips 2-251 - 1 maximum consecutive defects 2-251 - Cuts from: LS link 0 2-251 - Cuts from: LS link 1 2-268 - 2 defects found affecting 2 strips 2-268 - 1 maximum consecutive defects 2-268 - Cuts from: LS link 0 2-268 - Cuts from: LS link 1 2-285 - 5 defects found affecting 5 strips 2-285 - 1 maximum consecutive defects 2-285 - Cuts from: LS link 0 2-285 - Cuts from: LS link 1 2-302 - 3 defects found affecting 3 strips 2-302 - 1 maximum consecutive defects 2-302 - Cuts from: LS link 0 2-302 - Cuts from: LS link 1 2-319 - 3 defects found affecting 3 strips 2-319 - 1 maximum consecutive defects 2-319 - Cuts from: LS link 0 2-319 - Cuts from: LS link 1 2-336 - 3 defects found affecting 3 strips 2-336 - 1 maximum consecutive defects 2-336 - Cuts from: LS link 0 2-336 - Cuts from: LS link 1 2-353 - 4 defects found affecting 4 strips 2-353 - 1 maximum consecutive defects 2-353 - Cuts from: LS link 0 2-353 - Cuts from: LS link 1 2-370 - 2 defects found affecting 2 strips 2-370 - 1 maximum consecutive defects 2-370 - Cuts from: LS link 0 2-370 - Cuts from: LS link 1 2-388 - 4 defects found affecting 4 strips 2-388 - 1 maximum consecutive defects 2-388 - Cuts from: LS link 0 2-388 - Cuts from: LS link 1 2-404 - 3 defects found affecting 3 strips 2-404 - 1 maximum consecutive defects 2-404 - Cuts from: LS link 0 2-404 - Cuts from: LS link 1 2-421 - 3 defects found affecting 3 strips 2-421 - 1 maximum consecutive defects 2-421 - Cuts from: LS link 0 2-421 - Cuts from: LS link 1

Figures of gain and noise, and correlations with wafer/hybrid level test results, if they exist.