Module M4 results
The module name column is highlighted in GREEN once the component has succesfully reached BNL
Modules with stage BONDED are ready for thermal cycling:
Based on the stage FINISHED, the following 0 modules are ready to ship:
The box colours reflect the test status.
- GREEN - Test is present on the production database and passed.
- YELLOW - Test passed locally but was not yet uploaded.
- RED - Test failed. It may not have been uploaded.
As an extra check, the following text may be present in the boxes if the local test results are not consistent with the results from the database:
- ✓ DB ONLY: Test result is only found on the production database and not locally. This can happen if the test is uploaded directly to the database. Generally no action is necessary.
- ! DB MISSING: Test result cannot be found on the production database but is present locally. Please investigate and (re)upload.
- ! DB FAILED: Test result was failed on the production database but was not failed locally. Please investigate and (re)upload.
| Module | Stage | Flags | Post HV-tab IV | Gluing | Metrology | Wirebonding | TC Electrical | Shipped | Comment |
|---|---|---|---|---|---|---|---|---|---|
| SCIPP-M4-10 | |||||||||
| SCIPP-M4-09 | |||||||||
| SCIPP-M4-08 | |||||||||
| SCIPP-M4-07 | |||||||||
| SCIPP-M4-06 | |||||||||
| SCIPP-M4-05 | BONDED | ✓ DB ONLY | |||||||
| SCIPP-M4-04 | HV_TAB_ATTACHED | At some point during an IV scan of this sensor (post wirebonding). the corner of this sensor was chipped. The sensor had higher current than expected and earlier breakdown beyond 600V. The chip is near D2 and D3 on the frame. - Matthew Gignac, 05.01.2021--23-07-37
This module has its sensor rotated accidentally as shown in the slides here: https://indico.cern.ch/event/987604/contributions/415... |
|||||||
| SCIPP-M4-03 | HV_TAB_ATTACHED | This module has its sensor rotated accidentally as shown in the slides here: https://indico.cern.ch/event/987604/contributions/4157603/attachments/2165776/3655264/SCIPP%20Rotated%20Sensors%2018Dec2020.pdf - Matthew Gignac, 18.12.2020--18-57-18 | |||||||
| SCIPP-M4-02 | HV_TAB_ATTACHED | This module has its sensor rotated accidentally as shown in the slides here: https://indico.cern.ch/event/987604/contributions/4157603/attachments/2165776/3655264/SCIPP%20Rotated%20Sensors%2018Dec2020.pdf - Matthew Gignac, 18.12.2020--18-56-03 | |||||||
| SCIPP-M4-01 | GLUED | Start of test: HV: 300 V, 1.10 uA, LV: 11.0 V, 0.314 A, 5.0 % RH, 20.9 C. End of test: 1.22uA, 4.8%RH, 21.1C
- Vitaliy Fadeyev, 23.03.2020--19-45-20
Module metrology was performed with a slight variant of the v2a program. Had some issues with finding the edge of the sensor; so the first few steps were performed in edit mode to adjust lighting. On a second repeat, issues did not appear agai... |